Related Experiment Video
Updated: Jan 24, 2026

Author Spotlight: Developing a Point-of-Care Hemoglobin Estimation Method for Anemia Management
Published on: January 19, 2024
Continuous noninvasive hemoglobin monitoring estimates timing for detecting anemia better than clinicians: a
Bo Tang1, Xuerong Yu1, Li Xu1
1Department of Anesthesiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, Peking Union Medical College, Beijing, China; No.1 Shuaifuyuan, Wangfujing Street, Beijing, 100730, China.
Background:
Hemoglobin measurement is important for transfusion decision-making. Pulse CO-Oximetry provides real-time continuous hemoglobin (SpHb) monitoring. The triage role of SpHb trends based on hemoglobin measurements was investigated.
Methods:
In this diagnostic randomized controlled trial, 69 patients undergoing spine or cytoreductive surgery were randomly enrolled into SpHb-monitoring and standard-care groups. Diagnostic blood samples were drawn for CO-oximetry Hb (CoOxHb) when the SpHb decreased by 1 g/dl or at the clinician's discretion in the standard-care group. The positive predictive value (PPV) was defined as the ability to detect a decrease in CoOxHb > 1 g/dl or a CoOxHb < 10 g/dl; the PPVs were compared using Fisher's exact test. The SpHb and trend accuracies were calculated. The transfusion units and postoperative hemoglobin levels were compared.
Results:
The PPV of a decrease in CoOxHb > 1 g/dl was 93.3% in the SpHb group vs 54.5% without SpHb monitoring (p = 0.002). The PPV of CoOxHb < 10 g/dl was 86.7% vs. 50.0% for these groups (p = 0.015). The CoOxHb was never < 7 g/dl with SpHb monitoring. Sixty SpHb-CoOxHb data pairs and 28 delta pairs (ΔSpHb-ΔCoOxHb) were collected. The bias, precision and limits of agreement were - 0.29, 1.03 and - 2.30 to 1.72 g/dl, respectively. When ΔSpHb and ΔCoOxHb were > 1 g/dl, the concordance rate for changes in hemoglobin reached 100%. The delta pairs revealed a positive correlation [ΔSpHb = 0.49 * ΔCoOxHb - 0.13; r = 0.69, 95% confidence interval (0.53, 0.82)]. No significant differences were found in the transfusion volume or postoperative anemia state.
Conclusions:
The SpHb trend tracked changes in hemoglobin satisfactorily during surgery and more accurately estimated the appropriate timing for invasive hemoglobin measurements than the clinicians.
Trial Registration:
ChiCTR1800016290 (Prospective registered). Initial registration date was 24/05/2018.
More Related Videos
Related Concept Videos
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...
Continuous -time Fourier Transform
Basic Continuous Time Signals
The unit step function, denoted u(t), is zero for negative time values and one for positive time values, exhibiting a discontinuity at t=0. This function often represents abrupt changes, such as the step voltage introduced when turning a car's...
Sampling Continuous Time Signal
In the...
BIBO stability of continuous and discrete -time systems
To determine the BIBO stability, the convolution integral is utilized when a bounded continuous-time input is applied to a Linear Time-Invariant (LTI) system....
Time-Domain Interpretation of PD Control
Consider the example of control of motor torque. Initially, a positive...