Related Experiment Video
Updated: Jan 24, 2026

Author Spotlight: Innovative Methodology for Implanting and Securing Neural Probes in the Rodent Spinal Cord
Published on: July 12, 2024
On the development of equivalent medium for active implantable device radiofrequency safety assessment
Yu Wang1, Qingyan Wang1, Jingshen Liu1
1Department of Electrical and Computer Engineering, University of Houston, Houston, Texas.
Purpose:
To develop the equivalent medium theorem that can be used to perform accurate evaluation of implantable device safety under MRI exposure.
Methods:
Numerical methods were used to determine the equivalent medium parameters along clinically relevant trajectories inside a human body model. Additionally, numerical and experimental investigations were performed using both a computational human body model and an inhomogeneous phantom to demonstrate the effectiveness of the method.
Results:
The equivalent medium parameters, which are determined from a simplified lead configuration, are independent of the lead types and lead design parameters and only depend on the lead trajectories. Experimental investigations using an inhomogeneous phantom showed excellent agreement between the computational predicted values and the direct measured temperature rises indicating the effectiveness and accuracy of this method.
Conclusion:
For the models based on multiple patient trajectories studied, it demonstrates that the equivalent medium theorem is valid for leads of different types and designs, as long as the lead trajectories are determined.