Estimation of Nuwiq® (simoctocog alfa) activity using one-stage and chromogenic assays-Results from an international
Stefan Tiefenbacher1, Manuela Albisetti2, Peter Baker3
1Colorado Coagulation, Laboratory Corporation of America® Holdings, Englewood, Colorado.
Background:
Accurate determination of coagulation factor VIII activity (FVIII:C) is essential for effective and safe FVIII replacement therapy.
Fviii:
C can be measured by one-stage and chromogenic substrate assays (OSAs and CSAs, respectively); however, there is significant interlaboratory and interassay variability.
Aims:
This international comparative field study characterized the behaviour of OSAs and CSAs used in routine laboratory practice to measure the activity of Nuwiq® (human-cl rhFVIII, simoctocog alfa), a fourth-generation recombinant human FVIII produced in a human cell line.
Methods:
FVIII-deficient plasma was spiked with Nuwiq® or Advate® at 1, 5, 30 and 100 international units (IU)/dL. Participating laboratories analysed the samples using their routine procedures and equipment. Accuracy, inter- and intralaboratory variation, CSA:OSA ratio and the impact of different OSA and CSA reagents were assessed.
Results:
Forty-nine laboratories from 9 countries provided results. Mean absolute FVIII:C was comparable for both products at all concentrations with both OSA and CSA, with interproduct ratios (Nuwiq® :Advate® ) of 1.02-1.13. Mean recoveries ranged from 97% to 191% for Nuwiq® , and from 93% to 172% for Advate® , with higher recoveries at lower concentrations. Subgroup analyses by OSA and CSA reagents showed minor variations depending on reagents, but no marked differences between the two products. CSA:OSA ratios based on overall means ranged from 0.99 to 1.17 for Nuwiq® and from 1.01 to 1.17 for Advate® .
Conclusions:
Both OSAs and CSAs are suitable for the measurement of FVIII:C of Nuwiq® in routine laboratory practice, without the need for a product-specific reference standard.
More Related Videos
10:17Multi-target Chromogenic Whole-mount In Situ Hybridization for Comparing Gene Expression Domains in Drosophila Embryos
Published on: January 31, 2016
06:36Improved Enzyme Protection Assay to Study Staphylococcus aureus Internalization and Intracellular Efficacy of Antimicrobial Compounds
Published on: September 8, 2021
Related Concept Videos
Internal Receptors
Internal Energy
Internal Energy
What are Estimates?
The estimate for the mean of a sample is denoted by ͞x, whereas the mean of the population is designated as μ. Further, parameters such...
Comparing Mitochondrial, Chloroplast, and Prokaryotic Genomes
International System of Units
Prefixes are used to define both larger and smaller quantities in the SI system. For example, milli, micro, and nano define smaller quantities, while kilo, mega, and giga are used to define...