Katharine Klepin1, David Wing2, Michael Higgins2
1School of Exercise and Nutritional Sciences, San Diego State University, San Diego, CA.
Purpose:
Cardiorespiratory fitness (CRF), broadly defined as the body's ability to utilize oxygen, is a well-established prognostic marker of health, but it is not routinely measured. This may be due to the difficulty of acquiring high-quality CRF measures. The purpose of this study was to independently determine the validity of the Fitbit Charge 2's measure of CRF (Fitbit CRF).
Methods:
Sixty-five healthy adults between the ages of 18 and 45 yr (55% female, 45% male) were recruited to undergo gold standard V˙O2max testing and wear a Fitbit Charge 2 continuously for 1 wk during which they were instructed to complete a qualifying outdoor run to derive the Fitbit CRF (units: mL·kg·min). This measure was compared with V˙O2max measures (units: mL·kg·min) epoched at 15 and 60 s.
Results:
Bland-Altman analyses revealed that Fitbit CRF had a positive bias of 1.59 mL·kg·min compared with laboratory data epoched at 15 s and 0.30 mL·kg·min compared with data epoched at 60 s (n = 60). F statistics (2.09; 0.08) and P values (0.133; 0.926) from Bradley-Blackwood tests for the concordance of Fitbit CRF with 15- and 60-s laboratory data, respectively, support the null hypothesis of equal means and variances, indicating there is concordance between the two measures. Mean absolute percentage error was less than 10% for each comparison.
Conclusions:
The Fitbit Charge 2 provides an acceptable level of validity when measuring CRF in young, healthy, and fit adults who are able to run. Further research is required to determine if it is a potentially useful tool in clinical practice and epidemiological research to quantify, categorize, and longitudinally track risk for adverse outcomes.
Related Concept Videos
Reliability and Validity
Induced-fit Model
Enzymes exhibit substrate specificity, meaning that they can only bind to certain substrates. This is mainly determined by the shape and chemical...
Inclusive Fitness
Goodness-of-Fit Test
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Data Validation
Key parameters for method validation include: