Clinical Accuracy of the Nidek ARK-1 Autorefractor
Nabin Paudel1, Sameep Adhikari1, Ajit Thakur1
1Drishti Eye Care Centre, Kalanki, Kathmandu, Nepal.
Significance:
Autorefractors are commonly used by eye care practitioners worldwide as a starting point for clinical prescribing and by researchers as an instrument to study development of refractive errors and accommodation. This study demonstrates that the Nidek ARK-1 provides a reasonable and repeatable estimate of refractive error.
Purpose:
The purposes of this study were (a) to compare refraction measurements of the Nidek ARK-1 (2016 release) autorefractor with that of subjective refraction and retinoscopy performed by an experienced clinician and (b) to determine the intratest and intertest variability of autorefraction measures taken using the ARK-1 autorefractor.
Methods:
Sixty-seven adult patients aged 18 to 69 years underwent retinoscopy, subjective refraction, and ARK-1 autorefraction on a same day by a single clinician. A separate subset of 14 participants was invited for the repeatability and reproducibility study. Both eyes of each participant were included in the analysis.
Results:
A statistically significant (but not clinically significant) positive spherical difference was observed between the ARK-1 and subjective refraction (P = .003). Spherical equivalent refractive errors were statistically similar between the ARK-1 and subjective refraction (P = .20). A statistically and clinically significant difference was observed in the cylindrical component between the ARK and subjective refraction (P < .01). No statistically significant difference was observed between the ARK and subjective refraction in both the horizontal (J0; P = .08) and oblique cylindrical vector (J45; P = .96). Bland-Altman analysis revealed that the 95% limits of agreement were widest between the ARK and subjective refraction in all of the refractive components (-0.60 to 0.89 diopter for spherical component, -0.80 to 0.69 diopter for spherical equivalent, and -0.98 to 0.30 diopter for cylindrical component). The intertest and intratest variability of the ARK-1 was small.
Conclusions:
The Nidek ARK-1 autorefractor is a useful clinical tool that provides a reasonable and repeatable estimation of refractive error in adults.
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...