Comparison of Choroidal Thickness Measurements Using Spectral Domain Optical Coherence Tomography in Six Different
Helena Giannakaki-Zimmermann1,2, Wolfgang Huf3, Karen B Schaal1,2
1Department of Ophthalmology and Department of Clinical Research, Inselspital, Bern University Hospital, and University of Bern, Switzerland.
Purpose:
We investigate which spectral domain-optical coherence tomography (SD-OCT) setting is superior when measuring subfoveal choroidal thickness (CT) and compared results to an automated segmentation software.
Methods:
Thirty patients underwent enhanced depth imaging (EDI)-OCT. B-scans were extracted in six different settings (W+N = white background/normal contrast 9; W+H = white background/maximum contrast 16; B+N = black background/normal contrast 12; B+H = black background/maximum contrast 16; C+N = Color-encoded image on black background at predefined contrast of 9, and C+H = Color-encoded image on black background at high/maximal contrast of 16), resulting in 180 images. Subfoveal CT was manually measured by nine graders and by automated segmentation software. Intraclass correlation (ICC) was assessed.
Results:
ICC was higher in normal than in high contrast images, and better for achromatic black than for white background images. Achromatic images were better than color images. Highest ICC was achieved in B+N (ICC = 0.64), followed by B+H (ICC = 0.54), W+N, and W+H (ICC = 0.5 each). Weakest ICC was obtained with Spectral-color (ICC = 0.47). Mean manual CT versus mean computer estimated CT showed a correlation of r = 0.6 (P = 0.001).
Conclusion:
Black background with white image at normal contrast (B+N) seems the best setting to manually assess subfoveal CT. Automated assessment of CT seems to be a reliable tool for CT assessment.
Translational Relevance:
To define optimized OCT analysis settings to improve the evaluation of in vivo imaging.
More Related Videos
Related Concept Videos
Conservation of Protein Domains Over Different Proteins
A limited set of protein domains often duplicate and recombine during evolution. These domains can be organized in different combinations to...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Three-Domain System of Life
Membrane Domains
Protein Domains
The membrane comprises a group of distinct proteins responsible for carrying out a cell's specific function. For example, the plasma membrane of the human sperm, or a single germ cell, contains a unique set of proteins in the...
Three Developmental Domains
Physical Development
Physical processes, also known as maturation, encompass the biological changes that occur across an individual's life. These changes begin with genetic inheritance and continue through various stages, including growth in height and weight,...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...