Establishing how patient size and degree of miscentring affect CTDIvol, using patient data from a dose tracking
Sean Cournane1, Eimear Brunell2, Michael Rowan3
11 Department of Medical Physics and Bioengineering, St James's hospital , Dublin , Ireland.
Objectives:
The work aimed to investigate and model the relationship between CT dose index (CTDIvol), patient size and miscentring metrics using data extracted from a dose tracking software tool. Further, using the established model we aimed to extract and estimate further AEC metrics and, finally, demonstrate how the CTDIvol may vary across a range of imaging exams and with variations in patient size, positional metrics and with scan parameters.
Methods:
CT scan parameters, patient dose metrics, patient positioning information and patient and water equivalent diameter information was extracted from abdomen, thorax-abdomen-pelvis and thorax imaging exams acquired on a 128-slice Siemens Somatom Definition AS + CT system using dose tracking software over a 1 year period. A miscentring factor, accounting for the magnification of the patient due to miscentring was derived. Significant predictors ( p-value < 0.001) of the CTDIvol were considered as inputs into the model following regression analysis.
Results:
The model was capable of describing the CTDIvol and its variation with patient miscentring and patient size. AEC parameters, such as the reference CTDIvol, reference diameter and the AEC strength were estimated. Further, the model allowed for comparisons of how changes in scan settings, such as kVp, pitch and slice thickness affected CTDIvol.
Conclusions:
We demonstrate the use of clinical data, extracted from dose tracking software, to investigate and monitor AEC behaviour and CT output.
Advances In Knowledge:
The presented model allows for the remote investigation of AEC behaviour using patient data.
Related Concept Videos
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Dose Size and Dosing Frequency: Determination Methods
Patient-centered Care
One-Degree-of-Freedom System
A one-degree-of-freedom system is defined by an independent variable that determines its state and behavior. One example of a one-degree-of-freedom system is a simple harmonic oscillator, such as a...
Degrees of Freedom
For example, suppose there are three unknown numbers whose mean is 10; although we can freely assign values to the first and second numbers, the value of the last number can not be arbitrarily assigned.