Albuterol Delivery Efficiency in a Pediatric Model of Noninvasive Ventilation With a Single-Limb Circuit
Ariel Berlinski1, Jeanne Velasco2
1Department of Pediatrics, Pulmonology Section, University of Arkansas for Medical Sciences, and with the Pediatric Aerosol Research Laboratory, Arkansas Children's Research Institute, Little Rock, Arkansas. berlinskiariel@uams.edu.
Background:
Pediatric patients treated with noninvasive ventilation (NIV) are frequently given aerosol therapy. Limited pediatric data are available on the efficiency of aerosol delivery efficiency. We evaluated the effect of different nebulizers, positions in the single-limb ventilator circuit, and ventilator settings on the efficiency of aerosol delivery in a model of pediatric NIV. We hypothesized that using a vibrating mesh nebulizer, placing the nebulizer after the circuit leak, and not using the highest inspiratory positive airway pressure would increase aerosol delivery efficiency.
Methods:
We connected a breathing simulator in series to a low-dead-space filter holder (lung dose) and to an anatomically correct face/airway model of a 5-y-old child. A mask with an entrainment elbow was connected to a ventilator operated in a NIV bi-level mode and assembled with a single-limb heated-wired circuit. Inspiratory/expiratory pressures of either 15/5 or 20/5 cm H2O were used. We studied 3 different jet nebulizers and 2 vibrating mesh nebulizers loaded with albuterol solution (2.5 mg/3 mL). Albuterol was measured with spectrophotometry. The outcome measure was the efficiency of aerosol delivery (ie, lung dose expressed as percentage of the nominal dose).
Results:
Vibrating mesh nebulizers placed after the exhalation port of the circuit had the highest delivery efficiency, even compared with a vibrating mesh nebulizer integrated into the mask. Placing the nebulizer after the exhalation port of the circuit increased efficiency for all nebulizers. Vibrating mesh nebulizers were more efficient than jet nebulizers, regardless of their position in the circuit. Increasing the inspiratory pressure resulted in a variable effect on aerosol-delivery efficiency.
Conclusions:
In a model of pediatric NIV using a single-limb circuit, aerosol delivery devices were more efficient when placed after the exhalation port of the ventilator circuit. Vibrating mesh nebulizers were more efficient than jet nebulizers.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Power Distribution in Three-phase and Single Phase Circuits
Single-Phase Power Distribution:
Single-phase circuits are typical in household settings;...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Second-Order Circuits
Input signals typically originate from voltage or current sources, with the output often representing voltage across the capacitor and/or current through the inductor. For example, in...