Deep residual network for off-resonance artifact correction with application to pediatric body MRA with 3D cones
David Y Zeng1, Jamil Shaikh2, Signy Holmes2
1Department of Electrical Engineering, Stanford University, Stanford, California.
Purpose:
To enable rapid imaging with a scan time-efficient 3D cones trajectory with a deep-learning off-resonance artifact correction technique.
Methods:
A residual convolutional neural network to correct off-resonance artifacts (Off-ResNet) was trained with a prospective study of pediatric MRA exams. Each exam acquired a short readout scan (1.18 ms ± 0.38) and a long readout scan (3.35 ms ± 0.74) at 3 T. Short readout scans, with longer scan times but negligible off-resonance blurring, were used as reference images and augmented with additional off-resonance for supervised training examples. Long readout scans, with greater off-resonance artifacts but shorter scan time, were corrected by autofocus and Off-ResNet and compared with short readout scans by normalized RMS error, structural similarity index, and peak SNR. Scans were also compared by scoring on 8 anatomical features by two radiologists, using analysis of variance with post hoc Tukey's test and two one-sided t-tests. Reader agreement was determined with intraclass correlation.
Results:
The total scan time for long readout scans was on average 59.3% shorter than short readout scans. Images from Off-ResNet had superior normalized RMS error, structural similarity index, and peak SNR compared with uncorrected images across ±1 kHz off-resonance (P < .01). The proposed method had superior normalized RMS error over -677 Hz to +1 kHz and superior structural similarity index and peak SNR over ±1 kHz compared with autofocus (P < .01). Radiologic scoring demonstrated that long readout scans corrected with Off-ResNet were noninferior to short readout scans (P < .05).
Conclusion:
The proposed method can correct off-resonance artifacts from rapid long-readout 3D cones scans to a noninferior image quality compared with diagnostically standard short readout scans.