Comparison of routine and automated office blood pressure measurement
Robin Z Cheng1, Vivek Bhalla2, Tara I Chang2
1Stanford University School of Medicine.
Objectives:
From April to October 2018, we implemented a blood pressure measurement quality improvement project at our Hypertension Center. We aimed to compare blood pressure measured using routine, non-standardized office blood pressure and Systolic Blood Pressure Intervention Trial-like automated office blood pressure protocols.
Methods:
In 202 consecutive patients, we measured blood pressure using routine clinic methods and an automated office blood pressure protocol (5 min of rest followed by three blood pressure measurements at 1-min intervals).
Results:
The mean routine blood pressure was 145.6/76.4 mmHg and the mean automated office blood pressure was 135.3/70.1 mmHg. The mean paired difference in blood pressure was 10.3/6.3 mmHg, and Bland-Altman plots demonstrated wide limits of agreement. Using the systolic blood pressure goal of 130 mmHg, 26.9% of the patients not at goal by routine blood pressure were at goal by automated office blood pressure.
Conclusions:
Misclassifications of patient blood pressure control status and the wide variability between routine blood pressure and automated office blood pressure support the wider clinical implementation of automated office blood pressure to improve standardization, minimize incorrect blood pressure measurement and avoid over-treatment.
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer