A Modular Safety System for an Insulin Dose Recommender: A Feasibility Study
Chengyuan Liu1, Parizad Avari2, Yenny Leal3
1Centre for Bio-Inspired Technology, Department of Electrical and Electronic Engineering, Imperial College London, London, UK.
Background:
Delivering insulin in type 1 diabetes is a challenging, and potentially risky, activity; hence the importance of including safety measures as part of any insulin dosing or recommender system. This work presents and clinically evaluates a modular safety system that is part of an intelligent insulin dose recommender platform developed within the EU-funded PEPPER project.
Methods:
The proposed safety system is composed of four modules which use a novel glucose forecasting algorithm. These modules are predictive glucose alerts and alarms; a predictive low-glucose basal insulin suspension module; an advanced rescue carbohydrate recommender for resolving hypoglycemia; and a personalized safety constraint applied to insulin recommendations. The technical feasibility of the proposed safety system was evaluated in a pilot study including eight adult subjects with type 1 diabetes on multiple daily injections over a duration of six weeks. Glycemic control and safety system functioning were compared between the two-weeks run-in period and the end point at eight weeks. A standard insulin bolus calculator was employed to recommend insulin doses.
Results:
Overall, glycemic control improved over the evaluated period. In particular, percentage time in the hypoglycemia range (<3.0 mmol/l) significantly decreased from 0.82% (0.05-4.79) at run-in to 0.33% (0.00-0.93) at endpoint (P = .02). This was associated with a significant increase in percentage time in target range (3.9-10.0 mmol/l) from 52.8% (38.3-61.5) to 61.3% (47.5-71.7) (P = .03). There was also a reduction in number of carbohydrate recommendations.
Conclusion:
A safety system for an insulin dose recommender has been proven to be a viable solution to reduce the number of adverse events associated to glucose control in type 1 diabetes.
More Related Videos
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Bioavailability Study Design: Single Versus Multiple Dose Studies
Survey Safety
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Household Wiring And Electrical Safety