Related Experiment Video
Updated: Jan 24, 2026

Neuronavigation and Laparoscopy Guided Ventriculoperitoneal Shunt Insertion for the Treatment of Hydrocephalus
Published on: October 14, 2022
An approach using the occipital parietal point for placement of ventriculoperitoneal catheters in adults
Jason Duong1,2, Christopher J Elia1,2, Dan Miulli1,2
1Division of Neurosurgery, Riverside University Health System Medical Center, Moreno Valley, California, USA.
Background:
Ventriculoperitoneal shunts (VPS) have been widely used in the management of hydrocephalus. As current investigations into optimal approaches are being studied in the pediatric population, no general consensus on cranial entry points has been established for the adults. We compare conventional posterior and frontal approaches with an occipital parietal point (OPP) on computerized tomography (CT) while analyzing its associated outcomes.
Methods:
An Institutional Review Board (IRB) approved retrospective review was conducted on patients at a single institution between 1999 and 2016, with searches of CPT codes of 62223, 62230, 62258. The patient's lost to follow-up were excluded. Demographics, etiology of hydrocephalus, cranial entry points, and clinical outcomes (optimal placement, blood loss, operative time, malfunctions, or infections) were abstracted. Chi-square analyses were conducted to identify the association between treatment and clinical outcomes.
Results:
Ninety-three adults (≥18 years old) patients were included in the final analysis that had clinic follow-up, average age was 40.8 ± 15.6 years, with 57.0% had catheters placed utilizing the OPP, and 43.0% using conventional landmarks. OPP had less rates of suboptimal placement (P = 0.0469), and was less likely to develop a mechanical malfunction (5.7% vs. 12.5%). There was no difference in operative time, blood loss, or infection rate.
Conclusions:
Shunt malfunctions remain to be a common complication but can be reduced by optimal catheter positioning. The OPP established on computed tomography (CT) is just as safe as conventional landmarks, and can aid in optimal catheter positioning and can potentially reduce the risk of shunt malfunction secondary to suboptimal catheter placement.
Related Concept Videos
Adult Stem Cells
Frustration and Conflict: Approach-Approach, Approach-Avoidance
One common type of conflict is the Approach–Approach Conflict. In this case, a person faces two desirable...
Information Processing Approach
Actuarial Approach
Consider the example of a high-risk surgical procedure with significant early-stage mortality. A two-year clinical study is conducted,...
The Nativist Approach
Cardiopulmonary Resuscitation I: Adult