Evaluation of ocular pulse amplitude in patients with carotid artery stenosis using the Pascal dynamic contour
Gul Arikan1, Ziya Ayhan1, Mustafa Baris2
1Department of Ophthalmology, School of Medicine, Dokuz Eylul University, Izmir, Turkey.
Background:
To evaluate ocular pulse amplitude (OPA) in patients with carotid artery stenosis (CAS) using the Pascal dynamic contour tonometer (DCT).
Methods:
Consecutive patients who underwent ultrasound Doppler examination of the carotid artery for the suspicion of CAS were prospectively enrolled in the study. Each patient underwent a complete ocular examination. OPA was measured using Pascal DCT. The participants were divided into four groups according to the degree of CAS: Group 1 (no occlusion, control group), Group 2 (mild occlusion: < 50 per cent), Group 3 (moderate occlusion: 50-69 per cent), and Group 4 (severe occlusion: ≥ 70 per cent).
Results:
A total of 161 eyes of 81 patients were included in the study. Of the 81 patients, 50 (61.7 per cent) were male, and 31 (38.3 per cent) were female. The mean age was 66.6 ± 12.3 years (range, 20-91 years). Mean OPA values were found to be 2.68 ± 0.97 mmHg in Group 1 (n = 64 eyes), 2.62 ± 0.83 mmHg in Group 2 (n = 49 eyes), 2.30 ± 0.97 mmHg in Group 3 (n = 27 eyes) and 1.66 ± 0.59 mmHg in Group 4 (n = 21 eyes). There was no statistically significant difference in mean OPA levels between Group 1 and Group 2 (p = 0.73). However, statistically significant differences were found between the other groups when they were compared with each other (Groups 1 and 3, p = 0.02; Groups 1 and 4, p < 0.001; Groups 2 and 3, p = 0.02; Groups 2 and 4, p < 0.001; Groups 3 and 4, p = 0.002).
Conclusion:
OPA decreases in patients with moderate to severe CAS. OPA measurement with Pascal DCT can be used as a screening test for CAS.
More Related Videos
09:36A Magnetic Resonance Imaging-based Computational Protocol for Analysis of Plaque Morphology and Hemodynamics in Patients with Carotid Artery Stenosis
Published on: August 12, 2025
06:04Author Spotlight: Assessing the Cardiovascular Profile of Patients with Metabolic Syndrome
Published on: September 27, 2024
Related Concept Videos
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Pascal's Law
Application of Pascal's Law
Hydraulic systems are used to operate automotive brakes, hydraulic jacks, and numerous other mechanical systems. We can derive a relationship between the forces in a simple hydraulic system...
Topographic Surveying and Contours
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...