Yunchul Park1, Young Goun Jo, Kang-Ho Choi
1From the Division of Trauma, Department of Surgery (Y.P., Y.G.J.), Chonnam National University Hospital; Department of Neurology (K.-H.C.), Chonnam National University Hwasun Hospital; The Heart Research Center (M.K.), Chonnam National University Hospital Designated by the Ministry of Health, Welfare and Family Affairs of Korea; and Department of Nuclear Medicine (J.K.), Chonnam National University Hospital, Donggu, Gwangju, Korea.
Background:
Gamma probes can provide real-time information on lesion location and allow radioguided surgery in many diseases. The purpose of this study was to determine whether a gamma probe could be used to confirm balloon location in resuscitative endovascular balloon occlusion of the aorta and to determine the ease-of-use of the gamma probe localization method using human vasculature phantom and large animal.
Methods:
This investigation comprised a proof-of-concept study using human vasculature phantom and feasibility study using large animal. An assistant and two general surgeons with no prior experience in handling a gamma probe participated in both studies. A gamma probe was used by each surgeon to explore the location of a catheter balloon filled with 8 mL of normal saline containing 37 MBq of Technetium. The differences between predicted and actual balloon locations and the times taken to perform the trials were recorded. Balloon locations were confirmed by planar gamma camera images. The results were analyzed and compared according to the operators and experiments.
Results:
Overall sensitivity and specificity for the localization of the catheter balloon were 100% and 91% in phantom study, respectively, and 100% and 100% in animal study, respectively. The performance of the two operators for exploring the balloon location was similar in both the phantom and the animal studies. The mean time taken to explore the balloon locations with the aid of the gamma probe was only 2.36 minutes in phantom study and 2.65 minutes in the animal study.
Conclusion:
The gamma probe method was accurate and rapid for confirming balloon location. We suggest that the gamma probe has potential for confirming balloon placement in endovascular procedures without the need for fluoroscopy.
Related Concept Videos
Confirmation Biases
Labeling DNA Probes
Radioisotopes, fluorophores, or small molecule binding partners like biotin or digoxigenin, are the most widely used reporter tags for labeling DNA probes. These labels can be attached to the probe DNA molecule via...
Blind Procedures
Handwashing II: Pre-procedure and Initial Procedure Steps
The hand washing procedure itself includes the following steps. First, cover cuts, if any, on hands with a waterproof dressing. Cuts and abrasions can become contaminated with bacteria hindering the ability to clean the area thoroughly. In addition, repeated hand washing can worsen an injury. The nails must be short and clean, without nail...
Handwashing III: During the Procedure and Post-Procedure Steps
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...