Why low-voltage shock impedance measurements fail to reliably detect insulation breaches in transvenous
Charles D Swerdlow1, John E Porterfield2, Anil G Kottam2
1Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, California.
Background:
Implantable cardioverter-defibrillators (ICDs) use low-voltage measures of shock impedance (LVSZ) to monitor integrity of leads.
Objective:
To determine the separation distance between conductors required for LVSZ to detect insulation breaches that produce short circuits during shocks, causing failed defibrillation.
Methods:
We simulated in-pocket insulation breaches between the ICD generator (CAN) and cables to the distal coil of 10 leads from 2 manufacturers. The ICD and lead were placed in an electrolyte bath. Polystyrene sheets were used to control the breach-CAN separation. We determined both the maximum lead-CAN separation for shorts during 800 V shocks and the shock strength at which shorts occurred for a fixed separation. We also calculated breach impedance and measured it using a low-voltage instrument.
Results:
The maximum breach-CAN separation for shorting was 350-500 μm for all leads. The minimum shock strength to short varied from 650 to 771 V (24-32 J). LVSZ never triggered a warning, even with no separation between the cable's inner insulation and the CAN. Using low-voltage pulses, breach impedance was measured at approximately 500-1000 Ω.
Conclusion:
LVSZ is insensitive to insulation breaches that cause life-threatening, shorted shocks. The explanation likely relates to impedance differences between ionic conduction during LVSZ measurements and free-electron conduction in plasma discharges.
Related Concept Videos
Reliability and Validity
Insulation Coordination
Conductors and Insulators
Most metals are conductors. Their atomic configuration is such that one or more electron(s) are loosely bound to the nucleus in each atom. Thus, a sea of mobile electrons are available in them, known as free electrons. Their easy...
Hypothesis: Accept or Fail to Reject?
There are two ways to indicate that the null hypothesis is not rejected. 'Accept' the null...
Thermal Insulation in Masonry Walls
External insulation can be applied using an Exterior Insulation and Finish System (EIFS), which involves affixing panels of plastic foam to the wall and covering them with a polymeric stucco reinforced with glass fiber mesh....
Voltage