Method comparison study of the Elecsys® β-Amyloid (1-42) CSF assay versus comparator assays and LC-MS/MS
Leslie M Shaw1, Oskar Hansson2, Ekaterina Manuilova3
1Department of Pathology and Laboratory Medicine, Perelman School of Medicine, University of Pennsylvania, 3400 Civic Center Blvd, Philadelphia, PA 19104, USA.
Background:
Alzheimer's disease (AD) biomarkers, such as cerebrospinal fluid (CSF) amyloid-β (1-42; Aβ42), can provide high diagnostic accuracy. Several immunoassays are available for Aβ42 quantitation, but standardisation across assays remains an issue. We compared the Elecsys® β-Amyloid (1-42) CSF assay with three assays and two liquid chromatography tandem mass spectrometry (LC-MS/MS) methods.
Methods:
Three method comparison studies evaluated the correlation between the Elecsys® β-Amyloid (1-42) CSF assay versus: INNOTEST® β-AMYLOID(1-42) (860 samples) and the Roche Diagnostics-developed LC-MS/MS method (250 samples); INNO-BIA AlzBio3 and the University of Pennsylvania (UPenn)-developed LC-MS/MS method (250 samples); and ADx-EUROIMMUN Beta-Amyloid (1-42) enzyme-linked immunosorbent assay (ELISA) (49 samples).
Results:
High correlation was demonstrated between Elecsys® β-Amyloid (1-42) CSF and comparator assays: INNOTEST® β-AMYLOID(1-42) (Spearman's ρ, 0.954); INNO-BIA AlzBio3 (Spearman's ρ, 0.864); ADx-EUROIMMUN Beta-Amyloid (1-42) ELISA (Pearson's r, 0.925). Elecsys® assay and LC-MS/MS measurements were highly correlated: Pearson's r, 0.949 (Roche Diagnostics-developed method) and 0.943 (UPenn-developed method).
Conclusion:
Findings from this multicentre evaluation further support use of the Elecsys® β-Amyloid (1-42) CSF assay to aid AD diagnosis. CSF-based certified reference materials should improve agreement across assays and mass spectrometry-based methods, which is essential to establish a global uniform CSF Aβ42 cut-off to detect amyloid pathology.
Related Concept Videos
Amyloid Fibrils
Amyloid deposits were observed as early as 1639 in the liver and the spleen. In 1854, Rudolph Virchow performed iodine staining,...
Amyloid Fibrils
Antihypertensive Drugs: Action of β1 Blockers
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...