The influence of hematocrit on the hemodynamics of artificial heart valve using fluid-structure interaction analysis
Han Hung Yeh1, Oleksandr Barannyk2, Dana Grecov1
1Department of Mechanical Engineering, University of British Columbia, 2054-6250, Applied Science Lane, Vancouver, BC, V6T 1Z4, Canada; School of Biomedical Engineering, University of British Columbia, 2222 Health Sciences Mall, Vancouver, BC, V6T 1Z3, Canada.
Abstract:
Patients with severe aortic stenosis could regain the proper hemodynamic performance and cardiovascular output by restoring the diseased aortic valve with an artificial heart valve replacement. However, given the uniqueness of each patient, the hemodynamic improvements after an aortic valve replacement could vary. The biomechanical and hemodynamical parameters can be influenced by some major factors including the patient's blood pressure and hematocrit. Therefore, the objective of this study is to analyze the hemodynamics and valve mechanics of a bileaflet mechanical heart valve and investigate the hemorheological characteristics under the change of hematocrit. The fully coupled fluid-structure interaction (FSI) approach was used to model the hemodynamics and valve dynamics. Particle image velocimetry (PIV) experiments were conducted with in vitro benchtop model using ViVitro Pulse Duplicator to verify and validate the FSI models. The current numerical analysis revealed the hematocrit influenced the shear stress distributions over a cardiac cycle. The structural stresses in the mechanical valve were also affected by the distributions of the shear stress in the blood flow. Parameter dependencies found in the current study indicate that the hematocrit is influential when conducting patient-specific modelling of prosthetic heart valves.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Anatomy of the Heart
The Fluid Mosaic Model
Ionic Crystal Structures
Most monatomic ions behave as charged spheres, and their attraction for ions of opposite charge is the same in every direction. Consequently, stable structures for ionic compounds result (1) when ions of one charge are surrounded by as many ions as possible of the opposite...
Antibody Structure
Antibodies, also known as immunoglobulins (Ig), are essential players of the adaptive immune system. These antigen-binding proteins are produced by B cells and make up 20 percent of the total blood plasma by weight. In mammals, antibodies fall into five different classes, which each elicits a different biological response upon antigen binding.
The Y-Shaped Structure of Antibodies Consists of Four Polypeptide Chains
Antibodies consist of four polypeptide chains: two identical heavy...