Comparison of noise-magnitude and noise-texture across two generations of iterative reconstruction algorithms from
J Greffier1, A Larbi1, J Frandon1
1Department of Radiology, CHU Nîmes, University Montpellier, Medical Imaging Group Nîmes, 30029 Nîmes, France.
Purpose:
To compare the noise-magnitude and noise-texture across two generations of iterative reconstruction (IR) algorithms proposed by three manufacturers according to the dose level.
Materials And Methods:
Five computed tomography (CT) systems equipped with two generations of IR algorithms (hybrid/statistical IR [H/SIR] or full/partial model-based IR [MBIR]) were compared. Acquisitions on Catphan 600 phantom were performed at 120kV and three dose levels (3-, 7- and 12-mGy). Raw data were reconstructed using standard "soft tissue" kernel for filtered back projection and one iterative level of two generations of IR algorithms. Contrast to-noise-ratio (CNR) was computed using three regions of interest: two of them placed in the low-density polyethylene (LDPE) and Teflon® inserts and another placed on the solid water. Noise power spectrum (NPS) was computed to assess the noise-magnitude (NPS peak) and noise-texture (NPS spatial frequency).
Results:
CNR increased significantly in MBIR compared to H/SIR algorithms for General-Electric (GE) Healthcare (45%±12 [SD]) and Philips Healthcare systems (62%±11 [SD]) (P<0.001). Regarding Siemens Healthineers systems, CNR of MBIR was significantly lower than that of H/SIR (mean difference: -4%±5 [SD]) (P<0.001) for Teflon® insert but not for LDPE insert (mean difference: -4%±7 [SD]) (P=N.S.). NPS peaks were lower with MBIR than with H/SIR for GE Healthcare (-42%±8 [SD]) and Philips Healthcare (-75%±4 [SD]) systems, whereas it was greater with MBIR than with H/SIR for Siemens Healthineers (13%±11 [SD]) systems. NPS spatial frequencies were higher with MBIR than with H/SIR for Siemens (14%±10 [SD]) but lower for others (-17%±5 [SD] for GE Healthineers and -55%±3 [SD] for Philips Healthcare systems).
Conclusion:
This study demonstrates that recent MBIR algorithms, by comparison with the preceding generation, differ according to the main manufacturers with respect to noise-magnitude and noise-texture.
More Related Videos
Related Concept Videos
Orders of Magnitude
The order of magnitude is simply a way of rounding numbers consistently to the nearest power of 10. This makes doing rough mental math...
Shape and Texture of Coarse Aggregate
Trial and Error and Algorithm
Steel Manufacturing
During this smelting process, limestone plays a crucial role by forming slag. Slag captures impurities within the molten iron, such...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...