Assessing the concurrent validity of a gait analysis system integrated into a smart walker in older adults with gait
Christian Werner1,2, Georgia Chalvatzaki3, Xanthi S Papageorgiou3
1Centre for Geriatric Medicine, Heidelberg University, Heidelberg, Germany.
Objective:
To assess the concurrent validity of a smart walker-integrated gait analysis system with the GAITRite® system for measuring spatiotemporal gait parameters in potential users of the smart walker.
Design:
Criterion standard validation study.
Setting:
Research laboratory in a geriatric hospital.
Participants:
Twenty-five older adults (⩾65 years) with gait impairments (habitual rollator use and/or gait speed <0.6 m/s) and no severe cognitive impairment (Mini-Mental State Examination ⩾17).
Main Measures:
Stride, swing and stance time; stride length; and gait speed were simultaneously recorded using the smart walker-integrated gait analysis system and the GAITRite system while participants walked along a 7.8-m walkway with the smart walker. Concurrent criterion-related validity was assessed using the Bland-Altman method, percentage errors (acceptable if <30%), and intraclass correlation coefficients for consistency (ICC3,1) and absolute agreement (ICC2,1).
Results:
Bias for stride, swing and stance time ranged from -0.04 to 0.04 seconds, with acceptable percentage errors (8.7%-23.0%). Stride length and gait speed showed higher bias (meanbias (SD) = 0.20 (0.11) m; 0.19 (0.13) m/s) and not acceptable percentage errors (31.3%-42.3%). Limits of agreement were considerably narrower for temporal than for spatial-related gait parameters. All gait parameters showed good-to-excellent consistency (ICC3,1 = 0.72-0.97). Absolute agreement was good-to-excellent for temporal (ICC2,1 = 0.72-0.97) but only poor-to-fair for spatial-related gait parameters (ICC2,1 = 0.37-0.52).
Conclusion:
The smart walker-integrated gait analysis system has good concurrent validity with the GAITRite system for measuring temporal but not spatial-related gait parameters in potential end-users of the smart walker. Stride length and gait speed can be measured with good consistency, but with only limited absolute accuracy.
Related Concept Videos
Reliability and Validity
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Data Validation
Key parameters for method validation include:
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
The Integrated Rate Law: The Dependence of Concentration on Time
Formulating and Validating Nursing Diagnosis II
Risk nursing diagnoses represent clinical judgments of an individual, family, or community more vulnerable to developing the health problem than others...