Measuring Ascending Aortic Stiffness In Vivo in Mice Using Ultrasound
Published on: December 2, 2014
Noninvasive Estimation of Aortic Stiffness Through Different Approaches
Paolo Salvi1, Filippo Scalise2, Matteo Rovina3
1From the Department of Cardiovascular Neural and Metabolic Sciences, Istituto Auxologico Italiano, IRCCS, Milan, Italy (P.S., A.G., A.F., G.P.).
Noninvasive devices measuring carotid-femoral pulse wave velocity (PWV) accurately estimate aortic stiffness. Other methods, like finger-toe PWV, showed poor agreement with invasive measurements.
Area of Science:
- Cardiovascular Physiology
- Biomedical Engineering
- Clinical Diagnostics
Background:
- Aortic stiffness is a key indicator of cardiovascular health.
- Aortic pulse wave velocity (PWV) is a standard, noninvasive measure of aortic stiffness.
- Various noninvasive devices exist to estimate aortic PWV.
Purpose of the Study:
- To compare the accuracy of commonly used noninvasive devices for estimating aortic pulse wave velocity (PWV).
- To validate noninvasive aortic PWV measurements against invasive gold-standard methods.
- To identify reliable noninvasive devices for assessing aortic stiffness.
Main Methods:
- 102 patients undergoing coronary angiography had aortic PWV measured invasively.
- Seven noninvasive devices were used to estimate aortic PWV.
- Noninvasive measurements were compared with invasive aortic PWV and with each other.
Main Results:
- Carotid-femoral PWV devices (Complior Analyse, PulsePen ET/ETT, SphygmoCor) demonstrated strong agreement with invasive measurements (r > 0.64).
- Finger-toe PWV (pOpmètre) showed weak correlation with invasive and carotid-femoral PWV (r < 0.41).
- Devices using proprietary algorithms (BPLab, Mobil-O-Graph) showed weaker agreement and potential bias compared to carotid-femoral methods.
Conclusions:
- Noninvasive devices measuring carotid-femoral pulse wave velocity are reliable for assessing aortic stiffness.
- Finger-toe PWV and devices with proprietary algorithms showed limitations in accuracy.
- Clinical use of aortic PWV should prioritize validated carotid-femoral measurement techniques.
Related Concept Videos
What are Estimates?
The estimate for the mean of a sample is denoted by ͞x, whereas the mean of the population is designated as μ. Further, parameters such...
Estimation of k and VD of Aminoglycosides
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Estimation of the Physical Quantities
Aortic Regurgitation I: Introduction
Estimating Population Mean with Known Standard Deviation
The confidence interval estimate will have the form as follows:
(point estimate - error bound, point estimate +...