When deadman theory meets footprint decortication: a suture anchor biomechanical study
Chih-Kai Hong1, Kai-Lan Hsu1,2, Fa-Chuan Kuan1,2
1Department of Orthopaedic Surgery, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, No.138, Sheng-Li Road, Tainan City, 70428, Taiwan.
Background:
The optimal insertion angle for suture anchor insertion has long been of great interest. Although greater tuberosity decortication is commonly performed during rotator cuff repair, the effect of decortication on the suture anchor insertion angle remains unclear. The purpose of this study was to compare the pullout strength of threaded suture anchors inserted at 45° and 90° in decorticated and non-decorticated synthetic bone models.
Methods:
Two kinds of synthetic bones were used to simulate the decorticated and non-decorticated conditions, for which 40 metallic suture anchors were used. Anchors were inserted at 45° and 90° in both decorticated and non-decorticated models and tested under cyclic loading followed by load-to-failure testing. The number of completed cycles, ultimate failure load, and failure modes was recorded.
Results:
In the decorticated model, the ultimate failure load of anchors inserted at 45° (67.5 ± 5.3 N) was significantly lower than that of anchors inserted at 90° (114.1 ± 9.8 N) (p < 0.001). In the non-decorticated model, the ultimate failure load of anchors inserted at 45° (591.8 ± 58 N) was also significantly lower than that of anchors inserted at 90° (724.9 ± 94 N) (p = 0.003). Due to the diverse failure modes in the non-decorticated model, specimens with a failure mode of suture anchor pullout were analyzed in greater detail, with results showing a significantly larger pullout strength for anchors inserted at 90° (781.6 ± 53 N) than anchors inserted at 45° (648.0 ± 43 N) (p = 0.025).
Conclusion:
Regardless of decortication, the pullout strength of anchors inserted at 90° was greater than those inserted at 45°. The clinical relevance is that inserting suture anchors at 90° is recommended due to the significantly larger ultimate failure load in both decorticated and non-decorticated bones.
Related Concept Videos
Attribution Theory
The Anchoring-and-Adjustment Heuristic
Band Theory
The energy difference between these bands is known as the band gap.
Conductor, Semiconductor,...
Lipids as Anchors
The carboxy-terminal of most of the prenylated proteins, such as Ras proteins, contains...
Anchoring Junctions
Scientific Laws and Theories