Effectiveness of various decontamination methods on CAD/CAM camera mirror sleeves
Aim:
The purpose of this study was to evaluate the effectiveness of several methods of disinfection and sterilization of computer-aided design/computer-aided manufacturing (CAD/CAM) camera mirror sleeves (Omnicam; Dentsply Sirona).
Materials And Methods:
The outer surface of seven groups of mirror sleeves were inoculated by submerging them in suspensions of Staphylococcus aureus and Pseudomonas aeruginosa. Post inoculation, the groups were decontaminated as follows: Group A: no decontamination or sterilization following inoculation (positive control); Group B: surface cleaning with a neutral soap (Dawn Dish Soap, Procter & Gamble) and water only; Group C: surface disinfection with 17% isopropanol (CaviWipes; Metrex). Groups D to F received a different high-level disinfection (HLD) solution in an HLD container (Dentsply Sirona) as follows: Group D: 0.55% ortho-phthalaldehyde (OPA) (Cidex OPA; Johnson & Johnson); Group E: 7.5% hydrogen peroxide (H2O2) (Sporox II; Sultan); Group F: 7.35% H2O2 and 0.23% peracetic acid (PAA) (Compliance; Metrex). Group G received dry-heat sterilization (Rapid Heat Sterilizer; Cox). Also, dry-heat sterilized mirror sleeves that were not exposed to bacteria and not disinfected served as a negative control. The presence of bacteria was tested on the inside and outside of the sleeves by plating samples on TSA II. A percent reduction in CFU/ml from the positive control group was determined per group.
Results:
All methods of disinfection except Dawn Dish Soap resulted in greater than 99.99% reduction in CFU/ml compared with the positive control group.
Conclusions:
Both HLD or dry-heat sterilization resulted in no growth of microorganisms in cultures taken from both the outside and inside surfaces of the bacteria-contaminated mirror sleeves.
More Related Videos
03:10Pilot In Vitro Study to Assess Cleaning Ability and Effects of Different Decontamination Methods on Implant Surfaces
Published on: November 21, 2025
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
Related Concept Videos
C4 Pathway and CAM
C4 Pathway
The C4 pathway is used by plants such as...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Buffer Effectiveness
The buffer capacity is the amount of acid or base that can be added to a given volume...
Framing Effects
Biological Effects of Radiation
Self-Awareness and Its Effects