Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Method to quantitatively assess electrode migration from medical images: Feasibility and application in patients with
Eduardo Tanaka Massuda1, Thomas Demarcy2, Michel Hoen2
1a Department of Ophthalmology, Otorhinolaryngology, Head and Neck Surgery, Ribeirão Preto Medical School , University of São Paulo , Ribeirão Preto, São Paulo , Brazil.
Abstract:
Objective: To propose a method for quantitative assessment of the migration of lateral-wall, straight electrode arrays after surgery based on postoperative Cone Beam Computed Tomography (CBCT) images and automated medical image analysis techniques. Methods: A preliminary study is conducted on 19 implanted ears. For each implantation, two CBCT images are objectively analyzed. Electrode arrays are consistently projected into the same coordinate system in order to estimate precisely the migration of each electrode. Spatial configuration changes are characterized with the overall curvature of the electrode array. Results: From the samples analyzed no significant electrode migration, extrusion or electrode curvature changes were found. Mean infinitesimal local migration reveals a tendency where apical electrodes tend to move away from the modiolus and basal electrodes away from the round window. Conclusion: CBCT images demonstrate adequate resolution with limited artifacts to assess the electrode array position in vivo. Automated medical image analysis techniques and consistent coordinate system allow to quantitatively estimate migration and extrusion effect for lateral-wall, straight electrode array.
Related Concept Videos
Nephrotic Syndrome II : Assessment and Medical Management
Generation of Straight or Branched Actin Filaments
Arp2/3 Complex
Arp2/3 complex is a seven-subunit complex consisting of two proteins similar to actin- Arp2 and Arp3, and five other subunits that help keep Arp2 and Arp3 inactive. When required, the complex is...
Magnetic Field Due To A Thin Straight Wire
Magnetic Field Due to Two Straight Wires
Standard Electrode Potentials
Inhaled Medications