Contrast Extravasation using Power Injectors for Contrast-Enhanced Computed Tomography in Children: Frequency and
Christian A Barrera1, Ammie M White1, Ashley M Shepherd1
1Department of Radiology, The Children's Hospital of Philadelphia, 3401 Civic Center Boulevard, Philadelphia, PA 19104.
Purpose:
To evaluate the safety of power injectors for contrast-enhanced computed tomography (CT) in children, namely: the prevalence and injury severity of contrast extravasations related to power injectors and the factors associated with these events.
Methods:
The need to obtain informed consent was waived for this HIPAA-compliant and IRB approved retrospective study. Around 2429 contrast-enhanced CT performed with a power injector were identified during a 3-year period. Data collected included patient demographic, power injector, and contrast agent information. The patients' symptoms, severity of injury and treatment with contrast extravasation were recorded. Around 1496 cases (823 boys, 673 girls) were included in the analysis. Independent-sample t test and Chi-square were used. For a sub-analysis using the extravasation cases, nonparametric tests were used.
Results:
The mean age was 9.5 ± 6.1 years. The most common access site, catheter site, and contrast agent used were the antecubital fossa, 22 gauge and Iohexol. The mean peak pressure was 68.9 ± 62.3 psi and the flow rate was 1.7 ± 0.9 mL/s. Eighteen cases of contrast extravasation were identified with a mean age of 11.2 ± 6.2 years. There were seven mild, six moderate, and five severe. Cases with extravasation had significantly higher peak pressure (p < 0.001) and flow rate (p < 0.001) compared to those without extravasation. Patients who received Iohexol-350 had significantly more contrast extravasation compared to those who used Iohexol-300 (p = 0.03). However, after post-hoc correction, only peak pressure (p < 0.01) and flow rate (p = 0.01) remained significant.
Conclusion:
The use of power injectors in children undergoing contrast-enhanced CT is associated with a low rate of extravasation and of long-term injury.
More Related Videos
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
06:02Anatomical Reconstructions of the Human Cardiac Venous System using Contrast-computed Tomography of Perfusion-fixed Specimens
Published on: April 18, 2013
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Power
Frequency-dependent Selection
Power System Distribution
The transmission system is designed...