Evaluation of a Multiplex Fully Automated Treponemal and Nontreponemal (Rapid Plasma Reagin) Assay
Sophie Arbefeville1,2, Maureen Lynch2, Patricia Ferrieri1,2
1Department of Laboratory Medicine and Pathology, University of Minnesota Medical School, Minneapolis.
Objectives:
In June 2017, Bio-Rad Laboratories received US Food and Drug Administration clearance for its BioPlex 2200 Syphilis Total & RPR (rapid plasma reagin) assay. It is the first fully automated treponemal/nontreponemal multiplex flow immunoassay, simultaneously detecting Treponema pallidum and reagin antibodies and an RPR titer. We compared the performance of the BioPlex Syphilis Total & RPR assay with the LIAISON Treponema Assay and the manual BD Macro-Vue RPR 18-mm Circle Test.
Methods:
In total, 314 serum specimens were tested for treponemal immunoglobulin G/immunoglobulin M and RPR with the LIAISON Treponema Assay, the BioPlex 2200 Syphilis Total & RPR assay, and the manual BD Macro-Vue RPR card test. All discordant results were further tested with the T pallidum particle agglutination assay from Fujirebio Diagnostics.
Results:
The overall percent agreement for the BioPlex assay for treponemal antibodies with the LIAISON Treponema Assay was 96.1%. Sensitivity and specificity for the BioPlex RPR assay were 90.5% and 97.2%, respectively (the manual RPR assay was considered the gold standard).
Conclusions:
The BioPlex 2200 Syphilis Total & RPR assay performance was comparable to the LIAISON Treponema Assay and the manual RPR test. Compared with the manual RPR, the automation of RPR testing offered labor savings, objective result reporting, and improved workflow.
Related Concept Videos
Composition of Blood Plasma
Plasma Membrane in Bacteria and Archaea
Enlargement of the Plasma Membrane
Protein Buffers in Blood Plasma and Cells
Certain amino acids can exist in a zwitterion state at a...
Drug Distribution: Plasma Protein Binding
Fusion of Secretory Vesicles with the Plasma Membrane
In 1993, Jim Rothman proposed that the antiparallel pairing of vesicular and transmembrane SNAREs, or...