A Biomechanical Comparison of Four Hip Arthroplasty Designs in a Canine Model
Nathaniel R Ordway1, Kristian J Ash2, Mark A Miller1
1Department of Orthopedic Surgery, SUNY Upstate Medical University, Syracuse, New York, United States.
Objective:
The aim of this study was to develop an in vitro biomechanical protocol for canine cementless hip arthroplasty that represents physiological gait loading (compression and torsion) and to evaluate if three alternative implant designs improve fixation compared with the traditional collarless, tapered stem in the clinically challenging case of moderate canal flare index.
Study Design:
Twenty-four (six/group) laboratory-prepared canine constructs were tested using a simulated gait and overload (failure) protocol. Construct stiffness, failure load/displacement and migration were measured as outcome variables.
Results:
Simulated gait loading did not show any significant differences between implant types for peak displacement, peak rotation, torsional stiffness, subsidence or inducible displacement. The collared and collarless stem groups were stiffer in compression compared with the collarless with a lateral bolt and short-stem groups. Increasing the loading above simulated gait showed significant reductions in compressive and torsional stiffness for all implant constructs. Despite the reductions, the short-stem group showed significantly higher stiffness compared with the other three groups.
Conclusion:
Peak failure loads (compressive and torsional) in this study were approximately four to seven times the simulated gait loading (430 N, 1.6 Nm) regardless of implant type and highlight the importance of limiting activity level (trotting, jumping) following hip replacement in the postoperative period and during the osseointegration of the implant.
Related Concept Videos
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Group Design
Factorial Design
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Design Example: Creating a Hydraulic Model of a Dam Spillway