Related Experiment Video
Updated: Jan 24, 2026

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Simple calculation of the optimal insertion depth of esophageal temperature probes in children
Sang Hyun Hong1, Jaemin Lee1, Joon-Yong Jung2
1Department of Anesthesiology and Pain Medicine, Seoul St. Mary's Hospital, The Catholic University of Korea, Seoul, Republic of Korea.
Insights
Accurate core temperature monitoring in children requires optimal esophageal temperature probe (ETP) placement. This study developed a simple formula based on height to determine the correct ETP insertion depth in pediatric patients.
Area of Science:
- Pediatric Anesthesiology
- Critical Care Medicine
- Medical Instrumentation
Background:
- Accurate core body temperature monitoring is crucial in pediatric anesthesia and critical care.
- Esophageal temperature probes (ETPs) are vital for this, but optimal insertion depth in children lacks standardized guidance.
- Previous studies have not established a formula for calculating the ideal ETP insertion depth in pediatric patients.
Purpose of the Study:
- To develop a simple and reliable method for determining the optimal esophageal temperature probe (ETP) insertion depth in children.
- To establish a formula correlating ETP depth with anthropometric measurements in pediatric patients.
- To improve the accuracy and safety of core temperature monitoring in pediatric patients.
Main Methods:
- Retrospective analysis of 181 children (aged 3-13 years) undergoing pectus excavatum repair.
- Utilized preoperative CT scans, intraoperative X-rays, and actual ETP insertion depths to measure optimal placement.
- Linear regression analysis to correlate optimal ETP depth with patient age, weight, and height.
Main Results:
- Optimal ETP insertion depth showed the strongest correlation with patient height (R² = 0.920).
- The best-fit equation derived was 0.180 × height + 6.749 cm.
- Three simplified formulae were validated, with 'height/5 + 5 cm' identified as a practical and accurate predictor.
Conclusions:
- Pediatric esophageal temperature probe (ETP) insertion depth is closely related to patient height.
- A simple formula, 'height/5 + 5 cm', can reliably predict the optimal ETP insertion depth in children.
- This formula offers a practical tool for clinicians to ensure accurate core temperature monitoring in pediatric patients.
Abstract:
Placing an esophageal temperature probe (ETP) in the optimal esophageal site is important in various anesthetic and critical care settings to accurately monitor the core temperature of a pediatric patient. However, no reported study has provided a formula to calculate the optimal insertion depth of ETP placement in children based on direct measurement of the optimal depth. The aim of this study was to develop a simple and reliable method to determine the optimal depth of ETP placement in children via their mouth. Using preoperative chest computed tomography scans, intraoperative chest X-rays, and the actual depth of ETP insertion, we measured the optimal depth of ETP placement retrospectively in 181 children aged 3-13 years who underwent minimally invasive repairs of the pectus excavatum and removal of a pectus bar. A linear regression analysis was performed to assess the correlation of the optimal depth of ETP placement with the children's age, weight, and height. The optimal depth of ETP placement had a greater correlation with height than with age or weight, and the best-fit equation was '0.180 × height + 6.749 (cm) (R2 = 0.920).' We obtained three simplified formulae, which showed no statistically significant difference in predicting the optimal depth of ETP placement: height/6 + 8 (cm), height/5 + 4 (cm), and height/5 + 5 (cm). The optimal depth of ETP via children's mouths has a close correlation with height and can be calculated with a simple formula 'height/5 + 5 (cm)'.
Related Concept Videos
Calculating pH Changes in a Buffer Solution
Calculating the Equilibrium Constant
For example, gaseous nitrogen dioxide forms dinitrogen tetroxide according to this equation:
Esophageal Varices-I: Introduction
Calculating Standard Free Energy Changes
Numerical Calculations
The solution to a problem is obtained using different methods. While manually solving algebraic symbols is one of the most common methods, the graphical method is often preferred. Computers...
Net Torque Calculations