Daily Electrocardiogram Electrode Change and the Effect on Frequency of Nuisance Alarms
Katherine Shue McGuffin1, Sheri Ortiz
1Katherine Shue McGuffin, DNP, MSN, FNP-C, is the Doctor of Nursing Practice Program coordinator and clinical assistant professor at The University of North Carolina at Charlotte in Charlotte, North Carolina. She is also a family nurse practitioner with a background in cardiology currently working in Charlotte, North Carolina. Her research interests include quality improvement, doctoral education, cardiovascular nursing, preventive medicine, and early detection and treatment of skin cancer. She is a member of Sigma Theta Tau International, American Nurses' Association, North Carolina Nurses' Association, the American Association of Nurse Practitioners, the Dermatology Nurses' Association, and American Association of Colleges of Nursing. Sheri Ortiz, MSN, RN, ACNS-BC, CCRN, is a critical care clinical nurse specialist at Caromont Regional Medical Center in Gastonia, North Carolina. She is retired as a major after serving in the US Army Nurse Corps for 23 years. Positions included critical care/trauma nurse, instructor for the US Army Critical Care/Emergency Nursing Course, nurse manager, ROTC Brigade nurse counselor and recruiter, and current role as critical care clinical nurse specialist. Her research interests include cardiology nursing and critical care nursing. She is a member of the National Association of Clinical Nurse Specialists, American Association of Critical-Care Nurses, and Sigma Theta Tau.
Background:
In 2018, The Joint Commission identified false telemetry alarms as a significant technology hazard placing patients at risk of injury. Reasons include poor skin preparation when applying electrodes and improper placement of electrodes.
Objectives:
The purpose of this quality improvement project was to determine if changing electrocardiogram electrodes daily would decrease the frequency of nuisance alarms.
Methods:
Study design was quantitative/comparative on all patients receiving telemetry monitoring on a 36-bed adult inpatient cardiac telemetry unit. Data collection occurred for 14 days before the intervention and 14 days during the intervention of daily electrode change. Comparison analysis determined if frequency of alarms decreased after the intervention with daily electrode change.
Results:
Postintervention data showed a 74.15% reduction in telemetry alarms following implementation of a daily electrode change.
Discussion:
Daily electrocardiogram electrode changes may be an effective strategy for reducing nuisance alarms on telemetry units. Outcomes can be used in conjunction with existing evidence to drive current practice.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Standard Electrode Potentials
Classical Conditioning in Daily Life
John B. Watson and Rosalie Rayner famously demonstrated the development of fear through classical conditioning in their experiment with Little Albert. They paired the...
Frequency-dependent Selection
Global Climate Change