Biomechanical evaluation of osteoporotic fracture: Metal fixation versus absorbable fixation in Sawbones models
Xinbao Wu1, Ziyun Wang2, Hui Li3
1School of Clinical Medicine, Tsinghua University, China.
Objective:
The failure of osteoporotic fracture after internal fixation is mainly caused by the underlying bone loss and strength compromise. The aim of this study is to investigate whether absorbable internal fixation can provide adequate mechanical stability and a reduction in the incidence of failure of fixation caused by bone loss and stress shielding.
Methods:
A low density cancellous bone model was selected to compare the insertion of screw (screw-in), removal of screw (screw -out) and pull - out strength of absorbable screw and metal screw. The long bone model of thin cortical bone was used to create the transverse fracture model. The model was fixed with absorbable plate-screw system and metal plate-screw system respectively. The fatigue test and static bending test were compared. Moreover, the size of screw hole area was assessed.
Results:
The maximal screw - in and screw - out torque of the absorbable screw was significantly greater than that of the metal screw (P < 0.05), but there was no significant difference in pull-out test (P > 0.05). No visible failure occurred in fatigue test. There was no significant difference between the maximum load of static bending test (P > 0.05). The screw hole area of absorbable samples was significantly smaller than that of metal samples (P < 0.05).
Conclusions:
In this experimental set-up it was found that the stability of absorbable screws in osteoporotic bone was better than metal screws. The absorbable system tested can achieve good stability, and the destruction of osteoporotic bone is small, which can reduce the occurrence of bone failure. Considering that absorbable material avoids the need of second surgery (implant removal) and reduces the stress shielding effect, we believe that absorbable internal fixation can be considered for fixation treatment of osteoporotic fractures.
More Related Videos
05:45Author Spotlight: Exploring Regeneration in Axolotls Through Insights in Cellular and Molecular Mechanisms, Bone Healing, and Implications for Human Therapies
Published on: April 12, 2024
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
Related Concept Videos
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Carbon-dioxide Fixation
Bonding in Metals
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Alkali Metals
Table 1: Properties of the alkali metals
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...