Technical Note: Quality assurance and relative dosimetry testing of a 60 Co total body irradiator using optical
Irwin I Tendler1, Jeremy S Bredfeldt2,3, Rongxiao Zhang4
1Thayer School of Engineering, Dartmouth College, Hanover, NH, USA.
Purpose:
The aim of this study was to create an optical imaging-based system for quality assurance (QA) testing of a dedicated Co-60 total body irradiation (TBI) machine. Our goal is to streamline the QA process by minimizing the amount time necessary for tests such as verification of dose rate and field homogeneity.
Methods:
Plastic scintillating rods were placed directly on the patient treatment couch of a dedicated TBI 60 Co irradiator. A tripod-mounted intensified camera was placed directly adjacent to the couch. Images were acquired over a 30-s period once the cobalt source was fully exposed. Real-time image filtering was used; cumulative images were flatfield corrected as well as background and darkfield subtracted. Scintillators were used to measure light-radiation field correspondence, dose rate, field homogeneity, and symmetry. Dose rate effects were measured by modifying the height of the treatment couch and scintillator response was compared to ionization chamber (IC) measurements. Optically stimulated luminesce detector (OSLD) used as reference dosimeters during field symmetry and homogeneity testing.
Results:
The scintillator-based system accurately reported changes in dose rate. When comparing normalized output values for IC vs scintillators over a range of source-to-surface distances, a linear relationship (R2 = 0.99) was observed. Normalized scintillator signal matched OSLD measurements with <1.5% difference during field homogeneity and symmetry testing. Beam symmetry across both axes of the field was within 2%. The light field was found to correspond to 90 ± 3% of the isodose maximum along the longitudinal and latitudinal axis, respectively. Scintillator imaging output results using a single image stack requiring no postexposure processing (needed for OSLD) or repeat manual measurements (needed for IC).
Conclusion:
Imaging of scintillation light emission from plastic rods is a viable and efficient method for carrying out TBI 60 Co irradiator QA. We have shown that this technique can accurately measure field homogeneity, symmetry, light-radiation field correspondence, and dose rate effects.
More Related Videos
Related Concept Videos
Quality Assurance
Testing Water Quality
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Relative Risk
Quality of Water
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...