Robotic Instrument Failure-A Critical Analysis of Cause and Quality Improvement Strategies
Alexander Tapper1, Derek Leale1, Gregory Megahan2
1Department of Urology, Beaumont Health, Royal Oak, MI.
Objective:
To introduce a quality improvement initiative tracking robotic instrument failures on a per case basis. It is imperative to understand rates of failure, financial implications of failures, and identify factors suggesting common mechanisms of failure.
Materials And Methods:
Starting in January 1, 2014 a quality reporting system for failed robotic equipment began. Staff was instructed to submit an incident report when a robotic instrument failed and the instrument returned to central processing. Instruments were then returned to the manufacturer (Intuitive Surgical Inc, Sunnyvale, CA) for analysis and reimbursement. Results of failure analysis by the manufacturer, including reimbursement rates, were recorded and correlated with the procedure and surgical specialty.
Results:
A total of 3935 robotic cases were performed during the study period with a reported instrument failure incidence of 6.2% (247 total instruments). Etiology of instrument failure was as follows: tip or wrist (46.9%), cable (30.0%), unknown (12.6%), control housing (5.3%), and shaft (3.2%). Highest instrument failure incidence was seen in colorectal surgery cases at 4.0%, Urology had the lowest at 2.7%. Manufacturer reimbursement rate was 57.9%; the most common reason for denial being mishandling/misuse of equipment, determined by manufacturer analysis.
Conclusion:
Herein, we have demonstrated that improved process flow of reporting is necessary to better track incidence and etiology of instrument failures. Cost savings comes from improved training of not only surgeons but operating room and central processing staff in handling equipment to prevent high rates of reimbursement denial.
More Related Videos
05:47Simulation of a Scaled Assembly Process with Collaboration of a Robotic Arm and Monitoring through a Vision System for Quality Control
Published on: August 29, 2025
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Related Concept Videos
Instrument Transformers
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Critical Region, Critical Values and Significance Level
In hypothesis testing, a sample statistic is converted to a test statistic using z, t, or chi-square distribution. A critical region is an area under the curve in probability distributions demarcated by the critical value. When the test statistic falls in this region, it suggests that the null hypothesis must be rejected. As this region contains all those values of the...
Critical Values
Uncertainty in Measurement: Reading Instruments