An evaluation of three biological indicator systems in flash sterilization
1Infection Control Department, University of Massachusetts Medical Center, Worcester 01605.
Infection Control : IC
|August 1, 1987
Summary
The 3M Attest #1261 biological indicator (BI) demonstrated superior reliability in monitoring flash sterilization cycles compared to Amsco Proof Flash and Castle Tec Test. Attest #1261 effectively detected sterilization failures after a three-minute exposure, unlike the other tested BIs.
Area of Science:
- Sterilization and Infection Control
- Microbiology
- Biomedical Engineering
Background:
- Biological indicators (BIs) are critical for verifying sterilization efficacy.
- Flash sterilization requires specific BIs to ensure rapid and reliable monitoring.
- Assessing BI sensitivity and reliability is crucial for patient safety.
Purpose of the Study:
- To evaluate the sensitivity and reliability of two flash-sterilization-specific biological indicators (BIs) and one traditional spore strip.
- To compare the survive/kill ratios of 3M Attest #1261, Amsco Proof Flash, and Castle Tec Test.
- To determine the most satisfactory BI for monitoring three-minute flash sterilization cycles.
Main Methods:
- Tested BIs included 3M Attest #1261, Amsco Proof Flash, and Castle Tec Test.
- Exposure times ranged from zero (come-up time) to three minutes at 273°F in a gravity displacement sterilizer.
- Survival was assessed by media color change or turbidity after incubation at 55°C; cycles were replicated.
Main Results:
- Amsco Proof Flash exhibited technical difficulties, including vial crushing issues and media evaporation.
- Castle Tec Test showed insufficient resistance, with no survivors detected at any exposure time.
- 3M Attest #1261 demonstrated 100% survival at 0 and 1-minute exposures, 94% at 2 minutes, and 0% at 3 minutes.
Conclusions:
- 3M Attest #1261 was more satisfactory in monitoring three-minute flash sterilization cycles than Proof Flash and Tec Test.
- Proof Flash and Tec Test exhibited limitations in sensitivity and reliability for flash sterilization monitoring.
- Independent evaluation of BIs by healthcare institutions is recommended for optimal use.