Grating monochromator wavelength calibration using an echelle grating wavelength meter
Optics Express
|June 6, 2019
Summary
This study introduces a new method for calibrating monochromators using an echelle grating wavelength meter (WM). This approach allows for precise spectral calibration at flexible intervals, ensuring traceability to SI-standards for scientific instruments.
Area of Science:
- Optical Engineering
- Metrology
- Spectroscopy
Background:
- Monochromators require accurate spectral calibration for traceable characterization measurements.
- Current calibration methods using gas-discharge lamps or lasers have limitations in sampling flexibility.
Purpose of the Study:
- To present a novel method for traceable calibration of monochromator center wavelengths.
- To enable flexible sampling intervals and calibration outside the wavelength meter's sensitive range.
Main Methods:
- Utilizing an echelle grating wavelength meter (WM) for calibration.
- Demonstrating the calibration process on DLR's Calibration Home Base (CHB) and for the EnMAP mission.
Main Results:
- Achieved traceable calibration of monochromators at almost any sampling interval.
- Successfully calibrated monochromators outside the sensitive spectral range of the WM.
- Applied the method to calibrate instruments for imaging spectrometers and the EnMAP hyperspectral satellite mission.
Conclusions:
- The presented approach offers a flexible and accurate method for monochromator calibration.
- This technique enhances the traceability of spectral measurements for scientific applications.
- The method is suitable for calibrating critical instruments for Earth observation missions.
Related Concept Videos
The de Broglie Wavelength
33.1K
In the macroscopic world, objects that are large enough to be seen by the naked eye follow the rules of classical physics. A billiard ball moving on a table will behave like a particle; it will continue traveling in a straight line unless it collides with another ball, or it is acted on by some other force, such as friction. The ball has a well-defined position and velocity or well-defined momentum, p = mv, which is defined by mass m and velocity v at any given moment. This is the typical...
33.1K
Glassware Calibration
1.3K
Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
1.3K
Instrument Calibration
693
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
693
Plotting and Calibrating the Root Locus
453
Root loci often diverge as system poles shift from the real axis to the complex plane. Key points in this transition are the breakaway and break-in points, indicating where the root locus leaves and reenters the real axis. The branches of the root locus form an angle of 180/n degrees with the real axis, where n is the number of branches at a breakaway or break-in point.
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...
453
Calibration Curves: Correlation Coefficient
4.6K
In a linear calibration curve, there is a value called the calibration coefficient, denoted by 'r,' which measures the strength and the direction of association between two variables. The correlation coefficient value ranges from −1 to +1. A value of +1 indicates a perfect positive linear correlation, −1 denotes a perfect negative correlation, and 0 implies no correlation between the two variables. A positive correlation value establishes that as one variable increases, the...
4.6K
Calibration Curves: Linear Least Squares
4.2K
A calibration curve is a plot of the instrument's response against a series of known concentrations of a substance. This curve is used to set the instrument response levels, using the substance and its concentrations as standards. Alternatively, or additionally, an equation is fitted to the calibration curve plot and subsequently used to calculate the unknown concentrations of other samples reliably.
For data that follow a straight line, the standard method for fitting is the linear...
For data that follow a straight line, the standard method for fitting is the linear...
4.2K