Chlamydia trachomatis samples testing falsely negative in the Aptima Combo 2 test in Finland, 2019
Kaisu Rantakokko-Jalava1,2, Kati Hokynar3, Niina Hieta4,2
1Department of Clinical Microbiology, Turku University Hospital, Turku, Finland.
Abstract:
Since February 2019, over 160 Chlamydia trachomatis (CT) cases testing negative or equivocal by Aptima Combo 2 (AC2) but positive by Aptima CT test run with Panther instruments occurred in Finland. The AC2 test targets chlamydial 23S rRNA while the CT test targets 16S rRNA. Sequencing of 10 strains revealed a nucleotide substitution in 23S rRNA. The significance of this for the failure of the AC2 test to detect the variant is not yet known.