Quantification of Ethanedinitrile in Air Using a New and Accurate Gas Chromatography Method
Sam E Brierley1, Anthony R Adlam2, Matthew K D Hall3,4
1The New Zealand Institute for Plant & Food Research Limited, Private Bag 11600, Palmerston North 4442, New Zealand. sam.brierley@plantandfood.co.nz.
Abstract:
Compared to previously tested fumigants such as methyl bromide, sulfuryl fluoride and phosphine; ethanedinitrile (EDN) is a new fumigant which is being trialled around the world as a pre-plant soil treatment and as a quarantine and pre-shipment (QPS) treatment of commodities. To collect the data necessary to assess the effectiveness of this fumigant, an accurate analytical method is needed across a wide concentration range. We reviewed the methods of detection for EDN described in recently published fumigation studies and have developed and validated a method to quantify EDN in air using a gas chromatograph equipped with a flame ionization detector (GC-FID). Our tested method has a linearity, precision, accuracy, limit of detection (LOD) and limit of quantification (LOQ) of R2 0.9988, 1.36%, 98.8%, 0.750 ppm and 1.073 ppm, respectively. These values were determined using internationally recognised guidelines for the validation of non-standard analytical methods, which means that our method can be applied to the different validation requirements of regulatory agencies and countries. Our method can be used for experimental conditions that require detection at low and high concentrations simultaneously because it is accurate, fast (0.6 min) and repeatable across a concentration range of 1 to 40,000 ppm. This method will help to standardise the quantification of EDN by research groups and facilitate acceptance of data by regulatory organisations around the world.
More Related Videos
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....