Body Fluid Testing at John F. Kennedy Medical Center in Liberia
Callum P Swift1,2, Cozie Gwaikolo2, John Ssentamu3
1Tallaght University Hospital, Dublin, Ireland.
Objectives:
To apply a simple method to validate testing for albumin, glucose, lactate dehydrogenase (LDH) and total protein (TP) in peritoneal, pleural, and cerebrospinal fluids (CSF) at a hospital in Liberia.
Methods:
Serum and body fluid specimens were mixed to create 100% serum and 25%, 50%, 75%, and 100% fluid tubes, which were tested on a Biotecnica BT3500. Differences less than 10% between calculated and measured concentrations were considered acceptable.
Results:
The means (confidence intervals) of the percent differences were: albumin/peritoneal 12.8 (6.0-19.7), albumin/pleural 2.8 (1.3-4.2), albumin/CSF 4.8 (2.2-7.5), glucose/peritoneal 4.0 (1.9-6.0), glucose/pleural 4.4 (3.1-5.7), glucose/CSF 2.9 (1.8-4.0), LDH/peritoneal 9.5 (6.3-12.7), LDH/pleural 9.5 (5.4-13.6), LDH/CSF 9.2 (5.2-13.3), TP/peritoneal 7.6 (3.8-11.4), TP/pleural 3.8 (1.5-6.2), and TP/CSF 4.5 (1.0-8.1).
Conclusions:
All mean differences except for one were less than 10%, allowing for the adoption of clinical testing. The mixing study is a low-cost method for quality-assured testing that can be performed by resource-limited laboratories.
Related Concept Videos
Composition of Body Fluids
Body Water Content and Fluid Compartments
Lattice Centering and Coordination Number
Types of Unit Cells
Imagine taking a large number of identical...
Center of Gravity
Center of Gravity
To determine its location, the principle of moments can be utilized by dividing the object into...
Center of Mass
The knowledge of the center of mass can also help us to describe and predict the motion of objects. For example, when a ball is thrown...