Overranging and overbeaming measurement in area detector computed tomography: A method for simultaneous measurement
Atsushi Urikura1, Takanori Hara2, Tsukasa Yoshida3
1Division of Diagnostic Radiology, Shizuoka Cancer Center, Nagaizumi, Japan.
Purpose:
We propose a novel method to assess overbeaming and overranging, as well as the effect of reducing longitudinal exposure range, by using a dynamic z-collimator in area detector computed tomography.
Methods And Materials:
A 500-mm diameter cylindrical imaging plate was exposed by helical scanning in a dark room. The beam collimation of the helical acquisitions was set at 32 and 80 mm. Overbeaming and overranging with the dynamic z-collimator were measured.
Results:
The actual beam widths were approximately 39 and 88 mm at 32 and 80 mm collimation, respectively, and were relatively reduced owing to increased beam collimation. Overranging was 27.0 and 48.2 mm with a pitch of 0.83 and 1.49 at 32 mm collimation and 72.5 and 83.1 mm with a pitch of 0.87 and 0.99 at 80 mm collimation. The dynamic z-collimator relatively reduced the overranging by 17.3% and 17.1% for the 32 and 80 mm collimation, respectively.
Conclusion:
We devised a method to simultaneously measure overbeaming and overranging with only one helical acquisition. Although the dynamic z-collimator reduced the overranging by approximately 17%, wider collimation widths and higher pitch settings would increase the exposure dose outside the scan range.
More Related Videos
09:47Array Tomography Workflow for the Targeted Acquisition of Volume Information using Scanning Electron Microscopy
Published on: July 15, 2021
09:37Extracting Metrics for Three-dimensional Root Systems: Volume and Surface Analysis from In-soil X-ray Computed Tomography Data
Published on: April 26, 2016
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Measurement of Bioavailability: Pharmacokinetic Methods
Measurement of Bioavailability: Pharmacodynamic Methods
Measuring Reaction Rates
Measures of Central Tendency