Motion compensation in extremity cone-beam computed tomography
Alejandro Sisniega1, Gaurav K Thawait1,2, Delaram Shakoor2
1Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD, 21205, USA.
Objectives:
To evaluate the improvement in extremity cone-beam computed tomography (CBCT) image quality in datasets with motion artifact using a motion compensation method based on maximizing image sharpness.
Methods:
Following IRB approval, retrospective analysis of 308 CBCT scans of lower extremities was performed by a fellowship-trained musculoskeletal radiologist to identify images with moderate to severe motion artifact. Twenty-four scans of 22 patients (18 male, four female; mean, 32 years old, range, 21-74 years old) were chosen for inclusion. Sharp (bone) and smooth (soft tissue) reconstructions were processed using the motion compensation algorithm. Two experts rated visualization of trabecular bone, cortical bone, joint spaces, and tendon on a nine-level Likert scale with and without motion compensation (a total of 96 datasets). Visual grading characteristics (VGC) was used to quantitatively determine the difference in image quality following motion compensation. Intra-class correlation coefficient (ICC) was obtained to assess inter-observer agreement.
Results:
Motion-compensated images exhibited appreciable reduction in artifacts. The observer study demonstrated the associated improvement in diagnostic quality. The fraction of cases receiving scores better than "Fair" increased from less than 10% without compensation to 40-70% following compensation, depending on the task. The area under the VGC curve was 0.75 (tendon) to 0.85 (cortical bone), confirming preference for motion compensated images. ICC values showed excellent agreement between readers before (ICC range, 0.8-0.91) and after motion compensation (ICC range, 0.92-0.97).
Conclusions:
The motion compensation algorithm significantly improved the visualization of bone and soft tissue structures in extremity CBCT for cases exhibiting patient motion.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Dosage Compensation
In addition to sexual development, the X chromosome has genes involved in autosomal functions such as brain development and the immune system. Therefore, males and females with distinct numbers of X chromosomes will...
Compensation Mechanisms
Respiratory Compensation
This mechanism addresses metabolic-induced pH imbalances by adjusting breathing rates. Respiratory compensation begins within minutes of detecting a pH...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Absolute and Local Extreme Values