Dosimetric and preparation procedures for irradiating biological models with pulsed electron beam at ultra-high
Patrik Gonçalves Jorge1, Maud Jaccard2, Kristoffer Petersson1
1Institute of Radiation Physics, Lausanne University Hospital, Lausanne, Switzerland; Department of Radiation Oncology, Lausanne University Hospital, Lausanne, Switzerland; Radio-Oncology Laboratory, DO/CHUV, Lausanne University Hospital, Lausanne, Switzerland.
Purpose:
Preclinical studies using a new treatment modality called FLASH Radiotherapy (FLASH-RT) need a two-phase procedure to ensure minimal uncertainties in the delivered dose. The first phase requires a new investigation of the reference dosimetry lying outside the conventional metrology framework from national metrology institutes but necessary to obtain traceability, repeatability, and stability of irradiations. The second consists of performing special quality assurance procedure prior to irradiation.
Materials And Methods:
The Oriatron eRT6 (PMB-Alcen, France) is an experimental high dose-per-pulse linear accelerator, delivering a 6 MeV pulsed electron beam with mean dose-rates, ranging from a few Gy/min up to thousands of Gy/s. Absolute dosimetry is investigated with alanine, thermo-luminescent dosimeters (TLD) and radiochromic films as well as an ionization chamber for relative stability. The beam characteristic and dosimetry are prepared for three different setups.
Results:
A cross-check between alanine, films and TLD revealed a dose agreement within 3% for dose-rates between 0.078 Gy/s and 1050 Gy/s, showing that these dosimeters are suitable for absolute dosimetry for FLASH-RT. In absence of appropriate setup dependent corrections, active dosimetry can reveal dose deviations up to 15% of the prescribed dose. These differences reduce to less than 3% when our dosimetric procedure is applied.
Conclusion:
We developed procedures to accurately irradiate biological models. Our method is based on validated absolute dosimeters and extends their use to routine FLASH irradiations. We reached an agreement of 3% between the delivered and prescribed dose and developed the requirements needed for workflows of preclinical and clinical studies.
More Related Videos
Related Concept Videos
Increased pulse rate
Many factors can elevate the risk of developing tachycardia. These include advanced age, a family history of arrhythmias, and an...
Decreased pulse rate
There are specific risk factors that can elevate the likelihood of developing bradycardia. Advanced age is a significant factor, with...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Electron Orbital Model
The first shell is closest to the nucleus, and it has only one subshell with a single spherical orbital called the...
Preparation of Samples for Electron Microscopy
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...