Repeatability of choroidal thickness measurements with Spectralis OCT images
Jason K Lau1, Sin Wan Cheung1, Michael J Collins2
1School of Optometry, The Hong Kong Polytechnic University, Hung Hom, Hong Kong.
Objective:
To investigate the repeatability of choroidal thickness measurements determined from enhanced depth imaging optical coherence tomography (EDI-OCT) images of eyes after wearing single-vision spectacles (SV) and orthokeratology (ortho-k) lenses.
Methods And Analysis:
Two EDI-OCT images of 40 children (SV: 20, ortho-k: 20) taken at a single visit were analysed twice. Subjects in the ortho-k group had been wearing ortho-k for 1-4 weeks. The choroidal thickness was determined from each image using a graph theory-based software and, where appropriate, manual correction of choroidal boundaries was undertaken by an experienced examiner.
Results:
The mean (±SD) choroidal thickness was 227.3±42.2 µm for the SV subjects and 251.1±54.4 µm for the ortho-k subjects. The interimage differences in choroidal thickness were -0.99±3.54 and -1.14±5.03 µm for the SV and ortho-k subjects, respectively, and the limits of agreement were +5.96 to -7.93 and +8.72 to -11.00 µm, respectively.
Conclusion:
The coefficients of repeatability of choroidal thickness measurements from two EDI-OCT images taken at a single visit were 7.08 µm (SV) and 10.06 µm (ortho-k), suggesting that a change in choroidal thickness of less than 10 µm may not indicate a real change resulting from ortho-k lens wear.
More Related Videos
10:14Author Spotlight: Ex Vivo OCT-Based Multimodal Imaging of Human Donor Eyes for Research into Age-Related Macular Degeneration
Published on: May 26, 2023
05:51Author Spotlight: Unraveling the Impact of Mechanical Ventilation on Diaphragm Function and Patient Outcomes
Published on: November 3, 2023
Related Concept Videos
Bioequivalence Experimental Study Designs: Repeated Measures, Cross-Over, Carry-Over, and Latin Square Designs
Design Example: Deciding Thickness of Lubricating Fluid in a Shaft
To calculate the required thickness of the lubricant layer, the tangential velocity at the shaft's surface must first be determined. This velocity is calculated by converting the rotational speed to angular velocity...
Uncertainty in Measurement: Significant Figures
Uncertainty in Measurement: Accuracy and Precision
Measures of Central Tendency
Measurement: Standard Units