Sensitivity and specificity of an electronic nose in diagnosing pulmonary tuberculosis among patients with suspected
Antonia M I Saktiawati1,2,3, Ymkje Stienstra4, Yanri W Subronto1,3
1Department of Internal Medicine, Faculty of Medicine, Public Health and Nursing, Universitas Gadjah Mada, Yogyakarta, Indonesia.
Objective:
To investigate the potency of a hand-held point-of-care electronic-nose to diagnose pulmonary tuberculosis (PTB) among those suspected of PTB.
Methods:
Setting: Lung clinics and Dr. Sardjito Hospital, Yogyakarta, Indonesia. Participants: patients with suspected PTB and healthy controls. Sampling: 5 minutes exhaled breath. Sputum-smear-microscopy, culture, chest-radiography, and follow-up for 1.5-2.5 years, were used to classify patients with suspected PTB as active PTB, probably active PTB, probably no PTB, and no PTB. After building a breath model based on active PTB, no PTB, and healthy controls (Calibration phase), we validated the model in all patients with suspected PTB (Validation phase). In each variable (sex, age, Body Mass Index, co-morbidities, smoking status, consumption of alcohol, use of antibiotics, flu symptoms, stress, food and drink intake), one stratum's Receiver Operating Characteristic (ROC)-curve indicating sensitivity and specificity of the breath test was compared with another stratum's ROC-curve. Differences between Area-under-the-Curve between strata (p<0.05) indicated an association between the variable and sensitivity-specificity of the breath test. Statistical analysis was performed using STATA/SE 15.
Results:
Of 400 enrolled participants, 73 were excluded due to extra-pulmonary TB, incomplete data, previous TB, and cancer. Calibration phase involved 182 subjects, and the result was validated in 287 subjects. Sensitivity was 85% (95%CI: 75-92%) and 78% (95%CI: 70-85%), specificity was 55% (95%CI: 44-65%) and 42% (95%CI: 34-50%), in calibration and validation phases, respectively. Test sensitivity and specificity were lower in men.
Conclusion:
The electronic-nose showed modest sensitivity and low specificity among patients with suspected PTB. To improve the sensitivity, a larger calibration group needs to be involved. With its portable form, it could be used for TB screening in remote rural areas and health care settings.
More Related Videos
Related Concept Videos
Pulmonary Tuberculosis I
Causative Organism
The primary infectious agent causing tuberculosis is Mycobacterium tuberculosis, a slow-growing, acid-fast, aerobic rod that exhibits sensitivity to heat and ultraviolet light. Instances of Mycobacterium bovis and Mycobacterium avium contributing to the development of TB infection are rare.
Mode of...
Pulmonary Tuberculosis II
Here is a detailed explanation of its pathophysiology:
Transmission: The process begins when a person inhales droplet nuclei containing M. tuberculosis. These are typically released into the air when an individual with pulmonary or...
Pulmonary Tuberculosis V
Latent tuberculosis infection occurs when TB bacteria are present in a person's body, but are not causing illness or symptoms. It is not contagious, and preventive treatment is crucial to avoid the...
Pulmonary Tuberculosis III
The first classification is based on the development of the disease, and it includes the following categories:
Pulmonary Tuberculosis IV
Several diagnostic approaches are used to detect TB. The conventional method is the Tuberculin Skin Test (TST), also known as the Mantoux test. However, this method has...
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...