Background:
In the past, various simulation and measurement paradigms have been introduced and evaluated in order to improve frequency-specific measurement of the hearing threshold using early auditory evoked potentials (EAEP). A promising approach for improvement of detection of stimulus response is the usage of frequency-modulated chirp signals, which optimize the temporal synchrony of neuronal responses along a region of the basilar membrane.
Aim Of The Study:
This study validated the performance of three generated narrow-band chirp stimuli in combination with a level-adaptive simultaneous masker on a collective of normally hearing subjects.
Material And Methods:
In this study 25 normal hearing subjects took part after undergoing pure tone audiometry as well as an objective estimation of the auditory threshold using low, middle and high chirp stimuli. The characteristic EAEP parameters were visually identified before statistical analysis. The characteristic latency level function was conducted using measurements within a stimulus level range from 80 to 0 dB HL. Afterwards a comparison of objectively verified auditory threshold and subjective auditory threshold was conducted.
Results:
All objectively determined thresholds of the frequency-specific evoked EAEP were on average below 10 dB HL: low chirp at 8.2 dB HL, middle chirp at 5.8 dB HL and high chirp at 5.4 dB HL. The mean difference compared to subjectively determined auditory thresholds at all frequencies was below 3 dB and was not significant.
Conclusion:
Brainstem evoked response audiometry (BERA) using a band-limited and level-specific masked chirp stimulus is an efficient method for the determination of frequency-specific excitation thresholds in the clinical routine. The small, insignificant difference compared to the subjectively determined auditory thresholds makes usage of correction factors mostly redundant. Confirming the study results concerning low chirp stimuli so far, the low chirp BERA currently seems to be the method of choice for estimation of auditory threshold at low frequency ranges around 500 Hz.
Related Concept Videos
Hearing
Band Theory
The energy difference between these bands is known as the band gap.
Conductor, Semiconductor,...
Ratio Level of Measurement
A set of data measured using the ratio scale takes care of the ratio problem and provides complete information. Ratio scale data are like interval scale data, except they have a zero point and ratios can be calculated....
Ordinal Level of Measurement
Data measured using an ordinal scale are similar to nominal scale data, but there is one major difference. The ordinal scale data can be ordered. An example of ordinal scale data is a list of the top five national parks...
Interval Level of Measurement
Data measured using the interval scale are similar to ordinal level data because they have a definite arrangement. However, in the interval level of measurement, the differences between data values are meaningful even though the data does not have a starting point.
Temperature is measured using the interval scale. It is measurable data, and the difference between...
Nominal Level of Measurement
The data that cannot be measured but can be grouped into categories fall under the nominal level of measurement. Data that is measured using a nominal...