Simulation studies of time reversal-based protoacoustic reconstruction for range and dose verification in proton
Yajun Yu1, Zhongxing Li1, Dong Zhang1
1School of Physics and Technology, Wuhan University, Wuhan, 430072, China.
Purpose:
In vivo range verification in proton therapy is a critical step to help minimize range and dose uncertainty. We propose to employ a time reversal (TR)-based approach using proton-induced acoustics (protoacoustics) to reconstruct pressure/dose distribution in heterogeneous tissues.
Methods:
The dose distribution of mono-energetic proton pencil beam in a CT-based patient phantom was calculated by Monte Carlo simulation. K-wave toolbox was used to investigate protoacoustic pressurization, propagation and reconstruction in 2D. To address the tissue heterogeneity effect, a number of physical parameters, including mass density (ρ), speed of sound (c), volumetric thermal expansion coefficient (αV ), isobaric specific heat capacity (Cp ) and attenuation power law prefactor (α0 ), were empirically converted from CT number. The performance was evaluated using two figures of merit: mean square error (MSE) of pressure profiles and Bragg peak localization error (ΔBP ). The impact of six parameters of the TR inversion was examined, including number of sensors, sampling duration, sampling timestep, spill time, noise level and number of iterations.
Results:
The quantitative accuracy of TR reconstruction and its dependency on the selected parameters is presented. Under optimum conditions, the positioning accuracy of the Bragg peak can be controlled below 1 mm. For instance, MSE is 0.0123 and ΔBP is 0.59 mm under the following conditions (32 sensors, sampling duration: 600 µs, sampling timestep: 40 ns, spill time: 1 µs, no noise).
Conclusions:
The feasibility of TR-based protoacoustic reconstruction in 2D for proton range verification was first demonstrated. The approach is not only applicable to pencil beam, but also has potential to be extended to passive scattering systems.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
06:04Experimental Investigation of the Hierarchical Control in DC Microgrids Using a Real-time Simulator
Published on: February 14, 2025
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
Strategies of Self-Presentation II: Self-Verification
Bioavailability Study Design: Single Versus Multiple Dose Studies
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Dose Size and Dosing Frequency: Determination Methods