Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Comparison of two radiofrequency ablation devices for atrial fibrillation concomitant with a rheumatic valve
Qin Jiang1, Sheng-Zhong Liu1, Lu Jiang1
1Department of Cardiac Surgery, Sichuan Provincial People's Hospital, Affiliated Hospital of University of Electronic Science and Technology, Chengdu, Sichuan 610072, China.
Background:
Perioperative and median-term follow-up outcomes have not been compared among procedures using radiofrequency ablation devices for permanent atrial fibrillation with concomitant rheumatic valve disease. We compared the sinus rhythm restoration efficacy of "non-irrigation" ablation forceps and an "irrigation" ablation device in patients with rheumatic valve disease undergoing a modified Cox maze radiofrequency ablation procedure due to permanent atrial fibrillation.
Methods:
Data of 278 patients with rheumatic valve disease from the Cardiac Surgery Department of Sichuan Provincial People's Hospital who underwent the modified Cox maze radiofrequency ablation procedure between May 2013 and May 2017 were reviewed. The procedure was performed using "non-irrigation" ablation forceps (AtriCure, group A) in 149 patients and an "irrigation" ablation device (Medtronic, group M) in 129 patients. Data were collected prospectively, and follow-up was documented and compared between the groups.
Results:
The radiofrequency procedure duration was 28.9 ± 3.8 min in group A and 29.5 ± 2.8 min in group M (t = 1.623, P = 0.106). The predicted radiofrequency time to the left atrium diameter was (Ya = 0.4964 X + 0.3762, R = 0.74) in group A and (Ym = 0.4331 X + 4.3563, R = 0.8435) in group M. The sinus rhythm (SR) conversion rate without use of anti-arrhythmic drugs was similarly good in groups A and M, with 75.2%, 72.5%, and 70.5% vs. 73.6%, 71.3%, and 69.8% at discharge, 6 and 12 months, respectively (F = 0.084, F = 0.046, F = 0.046, P > 0.05, respectively).
Conclusion:
Two types of radiofrequency ablation devices characteristic of "non-irrigation" and "irrigation" bipolar ablation forceps were similarly efficient at SR restoration.
Related Concept Videos
Amyloid Fibrils
Amyloid deposits were observed as early as 1639 in the liver and the spleen. In 1854, Rudolph Virchow performed iodine staining,...
Amyloid Fibrils
Rheumatic Heart Disease I: Introduction
Rheumatic Heart Disease IV: Nursing Management
Fibril-associated Collagen
For example, the type II collagen fibrils in cartilage have covalently bound type IX fibril-associated collagens at regular intervals. Other types of fibril-associated collagens are...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...