Dosimetric evaluation of synthetic CT for head and neck radiotherapy generated by a patch-based three-dimensional
Anna M Dinkla1,2, Mateusz C Florkow3, Matteo Maspero1,2
1Department of Radiotherapy, Division of Imaging & Oncology, University Medical Center Utrecht, Utrecht, The Netherlands.
Purpose:
To develop and evaluate a patch-based convolutional neural network (CNN) to generate synthetic computed tomography (sCT) images for magnetic resonance (MR)-only workflow for radiotherapy of head and neck tumors. A patch-based deep learning method was chosen to improve robustness to abnormal anatomies caused by large tumors, surgical excisions, or dental artifacts. In this study, we evaluate whether the generated sCT images enable accurate MR-based dose calculations in the head and neck region.
Methods:
We conducted a retrospective study on 34 patients with head and neck cancer who underwent both CT and MR imaging for radiotherapy treatment planning. To generate the sCTs, a large field-of-view T2-weighted Turbo Spin Echo MR sequence was used from the clinical protocol for multiple types of head and neck tumors. To align images as well as possible on a voxel-wise level, CT scans were nonrigidly registered to the MR (CTreg ). The CNN was based on a U-net architecture and consisted of 14 layers with 3 × 3 × 3 filters. Patches of 48 × 48 × 48 were randomly extracted and fed into the training. sCTs were created for all patients using threefold cross validation. For each patient, the clinical CT-based treatment plan was recalculated on sCT using Monaco TPS (Elekta). We evaluated mean absolute error (MAE) and mean error (ME) within the body contours and dice scores in air and bone mask. Also, dose differences and gamma pass rates between CT- and sCT-based plans inside the body contours were calculated.
Results:
sCT generation took 4 min per patient. The MAE over the patient population of the sCT within the intersection of body contours was 75 ± 9 Hounsfield Units (HU) (±1 SD), and the ME was 9 ± 11 HU. Dice scores of the air and bone masks (CTreg vs sCT) were 0.79 ± 0.08 and 0.70 ± 0.07, respectively. Dosimetric analysis showed mean deviations of -0.03% ± 0.05% for dose within the body contours and -0.07% ± 0.22% inside the >90% dose volume. Dental artifacts obscuring the CT could be circumvented in the sCT by the CNN-based approach in combination with Turbo Spin Echo (TSE) magnetic resonance imaging (MRI) sequence that typically is less prone to susceptibility artifacts.
Conclusions:
The presented CNN generated sCTs from conventional MR images without adding scan time to the acquisition. Dosimetric evaluation suggests that dose calculations performed on the sCTs are accurate, and can therefore be used for MR-only radiotherapy treatment planning of the head and neck.
Related Concept Videos
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...
Long-patch Base Excision Repair
Convolution Properties II
The width property indicates that if the durations of input signals are T1 and T2, then the width of the output response equals the sum of both durations, irrespective of the shapes of the two functions. For instance, convolving two rectangular pulses with durations of 2 seconds and 1 second results in a function with a width of 3 seconds.
The area property asserts that the area under the...
Network Covalent Solids
To break or to melt a covalent network solid, covalent bonds must be broken. Because covalent bonds are relatively strong, covalent network solids are typically...
Convolution Properties I
The commutative property reveals that the input and the impulse response of an LTI (Linear Time-Invariant) system can be interchanged without affecting the output: