Comparison of Gingival Retraction Materials Using a New Gingival Sulcus Model
Marco Dederichs1, Mina D Fahmy2, Harald Kuepper1
1Policlinic of Prosthetic Dentistry and Material Science, Centre for Dental Medicine, Jena University Hospital, Jena, Germany.
Purpose:
To investigate the pressure generated by different retraction materials using a novel gingival sulcus model.
Materials And Methods:
A gingival sulcus model was made using a polymer frame filled with silicon. A pressure sensor and a sulcus-fluid simulation were embedded into the silicon chamber to evaluate the pressure generated by different retraction materials. Six sizes of Ultrapak retraction cords (Ultradent, sizes #000 - 3), 4 retraction pastes (Expazen, Expasyl, Acteon, Access Edge, Traxodent) and 2 retraction gels (Sulcus Blue, Racegel) were analyzed. The mean and median pressure, interquartile range, and standard deviation (SD) of n = 10 repeated measurements were calculated. Statistical analysis was conducted by Kruskal-Wallis test for differences between the main groups of retraction materials, and Mann-Whitney U-test was performed to analyze differences between the single retraction materials.
Results:
Pressure (mean ± SD) generated by retraction cords increased with increasing size (48.26 ± 11.29 kPa, size #000 to 149.27 ± 28.75 kPa for #3). There was a significant difference between sizes (p < 0.01), except in #0 versus #1, and #2 versus #3. Retraction pastes generated pressures that ranged from 82.74 ± 29.29 kPa (Traxodent) to 524.35 ± 113.88 kPa (Expasyl). Retraction gels generated pressures from 38.96 ± 14.68 kPa (Racegel) to 95.15 ± 24.18 kPa (Sulcus Blue). Pressure generated by Expasyl was significantly higher than pressure generated by all other tested materials (p < 0.001).
Conclusion:
Pressure generated by retraction pastes and gels depends on the consistency of the retraction material, while pressure generated by retraction cords increased with increasing size of cords. Expasyl was found to generate the highest pressure compared to all other retraction materials.
Related Concept Videos
Clot Retraction and Fibrinolysis
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Members Made of Elastoplastic Material
As the bending moment...