Arrhythmic episodes in patients implanted with a cardioverter-defibrillator - results from the Prospective Study on
François Regoli1, Denis Graf2, Beat Schaer3
1Department of Cardiology, Fondazione Cardiocentro Ticino, Via Tesserete 48, 6900, Lugano, Switzerland. francois.regoli@cardiocentro.org.
Background:
Little is known about the ICD performance using enhanced detection algorithms in unselected, non-trial patients. Performance of recent generation ICD equipped with SmartShock™ technology (SST) for detection and conversion of ventricular tachyarrhythmias (VTA) was investigated.
Methods:
4P was a prospective, multicenter, observational study conducted in 10 Swiss implanting centers. Patients with a Class I indication according to international guidelines were included and received an ICD with SST. ICD discrimination capability was assessed by evaluating SST performance; therapy efficacy was assessed by rate of VTA conversions by ATP and by rescue shocks.
Results:
Overall, 196 patients were included in the analysis with a mean duration of follow-up of 27.7 months (452 patient-years of observation). Patient-specific rather than recommended programming was preferred. Device-detected episodes were frequent (5147 episodes in 146 patients, 74.5%). In 44 patients (22.4%), 1274 episodes were categorized as VTA; only 215 episodes were symptomatic. ATP was the first-line therapy and highly effective (99.9% success rate at the episode level, 100.0% at the patient level). Rescue shocks were rare (66 episodes in 28 patients); 7 shocks in 5 patients (2.6%) were inappropriate. Death and hospitalization rates were low.
Conclusions:
In a cohort of non-trial, unselected ICD patients, VTA episodes were frequent. The 4P results confirm the robustness of VTA detection by SST and the effectiveness of ATP treatment, hence limiting overall ICD shock burden.
Related Concept Videos
ATP Yield
The ETC is embedded in the inner mitochondrial membrane and is comprised of four main protein complexes and an ATP synthase. NADH and FADH2 pass electrons to these complexes, which pump protons into the intermembrane space. This distribution of...
Hydrolysis of ATP
If one phosphate group is removed, a molecule of ADP—adenosine diphosphate—remains, along with inorganic phosphate. ADP can be further hydrolyzed to AMP—adenosine...
Predicting Molecular Geometry
ATP Synthase: Structure
ATP and Macromolecule Synthesis
Most macromolecules are composed of single subunits, or building blocks, called monomers. The monomers combine with each other using covalent bonds to form larger molecules known as polymers.
Conversion of...
ATP and Energy Production