Design and operation verification of an automated pressure mapping and modulating seat cushion for pressure ulcer
Wei Carrigan1, Pavan Nuthi1, Charu Pande1
1University of Texas at Arlington Research Institute, Fort Worth, TX, USA.
Abstract:
A sensorized air cell-based seat cushion system was developed to address the issues of loading magnitude and duration at a sitting interface to aid in reducing risk of sitting acquired pressure ulcers. This system is capable of pressure mapping, redistribution, and offloading which were verified using an anthropomorphic model and a human subject. The system is comprised of an air cell array cushion, a pneumatic control unit, and a graphical user interface. ISO load deflection testing confirmed that the cushion's loading response is comparable to commercial air cell-based seat cushions. Testing demonstrated that the internal pressure of the air cells are indicative of interface pressure and can be used as input to pressure modulating algorithms. Uniform pressure distribution was achieved through automated pressure redistribution algorithm implementation where the immersion of a subject into the seat cushion increased and interface pressure decreased. High pressure point identification and automatic offloading were performed in which newly created high pressure points were addressed using subsequent redistribution. Pressure mapping enabled offloading and redistribution can objectively manage the effects of loading magnitude and duration at the sitting interface.
Related Concept Videos
Vapor Pressure
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Constant Pressure Calorimetry
Vapor Pressure Lowering
Osmosis and Osmotic Pressure of Solutions
Pressure Gauges