Laparoscopic Essure® Device Removal by En Bloc Salpingectomy-Cornuectomy with Intraoperative X-Ray Checking: A
Laura Miquel1, Patrice Crochet2, Sarah Francini2
1Department of Gynecology, Obstetrics, and Reproductive Medicine, Pôle femmes parents enfants, AP-HM La Conception University Hospital, Marseille, France (Drs. Miquel, Boubli, and Agostini).
Study Objective:
To evaluate the feasibility, efficacy, and morbidity of Essure® device removal via laparoscopic en bloc salpingectomy-cornuectomy resection (LSC) and the utility of confirmation of complete removal with an intraoperative X-ray of the removed specimen (IX-S).
Design:
Retrospective observational cohort study.
Setting:
Academic hospitals of La Conception, Marseille, and Arnaud-de-Villeneuve, Montpellier, France.
Patients:
Women who were not seeking future fertility seeking Essure® device removal by salpingectomy, between February 2017 and August 2018.
Interventions:
All women underwent LSC. An IX-S was performed to confirm complete Essure® device removal. In the case of an unsatisfactory IX-S, an intraoperative pelvic X-ray control of the patient (IX-P) was performed. If IX-P diagnosed residual Essure® fragments, an additional resection was performed, and the removed tissue was checked by IX-S.
Measurements And Main Results:
We included 72 women, and a total of 140 Essure® devices were removed. The IX-S confirmed complete Essure® device removal in 131 of 140 cases (93.6%) in 63 of 72 women (87.5%). Out of the 9 women with unsatisfactory IX-S, 6 had no residual Essure® fragments at IX-P, and Essure® device removal was considered complete. Three women had a persistent Essure® fragment at IX-P: an additional resection allowed complete removal in 2 cases and resulted in failure in 1 of 140 case (0.7%). There were 2 of 72 women (2.8%) intraoperative complications and 4 of 72 women (5.6%) postoperative grade 1 complications according to the Clavien-Dindo classification.
Conclusion:
Essure® device removal by LSC appears to be an effective and safe procedure. IX-S is a useful method to evaluate whether the removal of Essure® device is complete during an LSC procedure.
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Protein Folding Quality Check in the RER
X-ray Imaging
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
Radiological Investigation I: X-ray and CT