Related Experiment Video
Updated: Jan 23, 2026

A Volumetric Method for Quantification of Cerebral Vasospasm in a Murine Model of Subarachnoid Hemorrhage
Published on: July 28, 2018
Validation of MRI for Volumetric Quantification of Atelectasis in the Perioperative Period: An Experimental Study in
Eric Noll1,2,3, Mickael Ohana4, Maryse Hengen1,2
1Institut Hospitalo-Universitaire "Image-Guided Surgery", Université de Strasbourg, Strasbourg, France.
Abstract:
Background: Impairment of pulmonary aeration is a frequent postoperative complication that is associated with adverse outcome. Diagnosis and quantification of impaired pulmonary aeration by CT scan is limited due to concern for exposure to ionizing radiation. Magnetic resonance imaging (MRI) represents a potential radiation-free alternative for this use. We undertook an experimental study to validate the use of MRI to quantify pulmonary aeration impairment. Methods: Ten large white pigs were studied before intubation, after intubation, 2 h after non-protective mechanical ventilation and after intra-tracheal negative pressure suction to induce atelectasis. A lung CT scan immediately followed by a lung MRI were performed at all four time points. On the 40 CT images lung volumes corresponding to non-aerated, poorly aerated, normally aerated, and overinflated voxels were measured based on their radiodensity. Similarly, on the 40 MRI images lung volumes corresponding to non-aerated and aerated voxels were measured based on their signal intensity. The correlation between non-aerated lung by MRI vs., CT scans, and with PaO2/FiO2 measured at each of the four time points was assessed with the Pearson' correlation coefficient, bias and limits of agreement. Results: Pearson correlation coefficient, bias and limits of agreements between the CT non-aerated lung volumes and MRI abnormal lung volumes were 0.88, -16 ml, and (-108, 77), respectively. Pearson correlation coefficient between PaO2/FiO2 and abnormal lung volumes measured with MRI was -0.60. Conclusion: In a preclinical swine model, quantitative measurements of pulmonary atelectasis by MRI-imaging are well correlated with the gold standard, i.e., densitometric scan CT measurements.
Insights
Magnetic resonance imaging (MRI) can accurately quantify pulmonary aeration impairment, offering a radiation-free alternative to CT scans. This study validates MRI
Area of Science:
- Pulmonary medicine
- Radiology
- Medical imaging
Background:
- Pulmonary aeration impairment is a common and serious postoperative complication.
- Computed tomography (CT) scans are limited by radiation exposure concerns.
- Magnetic resonance imaging (MRI) offers a radiation-free alternative for assessing lung aeration.
Purpose of the Study:
- To validate the use of MRI for quantifying pulmonary aeration impairment.
- To compare MRI-based measurements with CT scan data.
- To assess the correlation between MRI findings and blood oxygenation levels.
Main Methods:
- A preclinical study involving ten pigs was conducted.
- Lung CT scans and MRI were performed at four time points, including induced atelectasis.
- Quantitative analysis of lung volumes (non-aerated, poorly aerated, normally aerated) was performed using both imaging modalities.
Main Results:
- MRI-derived abnormal lung volumes showed a strong correlation (Pearson's r=0.88) with CT-derived non-aerated lung volumes.
- Bias and limits of agreement between MRI and CT were -16 ml and (-108, 77) ml, respectively.
- Abnormal lung volumes measured by MRI correlated moderately with PaO2/FiO2 (r=-0.60).
Conclusions:
- Quantitative pulmonary atelectasis measurement using MRI is well-correlated with CT scan measurements in a preclinical swine model.
- MRI provides a reliable, radiation-free method for assessing pulmonary aeration impairment.
- This validates MRI as a promising tool for postoperative lung monitoring.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
The Periodic Table
Periodic Classification of the Elements
Reliability and Validity
Effective Value of a Periodic Waveform
The effective value of a periodic current represents the direct current (DC) that conveys the same average power to a resistor as the periodic current itself. This concept is crucial when assessing AC circuits. To determine the...
The Periodic Table and Organismal Elements